Investigation of optimal placement of locking attachment plate in periprosthetic femoral fractures : a biomechanical comparative study.
Koki Omori, Tomonori Baba, Michael Blauth, Yukie Horikoshi, Yuki Nagata, Muneaki Ishijima
Department of Orthopaedics, Juntendo University Faculty of Medicine, Tokyo, Japan. · Department for Orthopaedics and Traumatology, Medical University of Innsbruck, Innsbruck, Austria.
Key points
From the paper's conclusion- This biomechanical study demonstrates that proximal and dual LAP configurations provide higher static failure strength than diaphyseal LAP or LCP-only constructs, while resistance to early cyclic fracture was similar among configurations. The absence of early fatigue failure in the dual LAP configuration indicates a mechanical advantage when anatomical placement allows.
Abstract
AIMS: To investigate the biomechanical effects of proximal, diaphyseal, and dual locking attachment plate (LAP) configurations compared with a locking compression plate (LCP)-only construct in a Vancouver type C periprosthetic femoral fracture model under cyclic and monotonic loading. METHODS: A cementless stem was implanted into synthetic femora to create a standardized Vancouver type C periprosthetic femoral fracture. Four constructs (n = 6 each) were evaluated: proximal LAP, diaphyseal LAP, dual LAP, and a LCP-only control. Axial cyclic loading was applied starting at 600 N with progressive increments up to 1,800 N (maximum 6,500 cycles), followed by monotonic loading to failure. Primary outcomes were cyclic load to first fracture line and monotonic failure load. Composite bone distance and fracture line locations were also assessed. RESULTS: Cyclic load to first fracture line did not differ among constructs (all p > 0.05). Under monotonic loading, proximal LAP (mean 4,850.2 N (SD 1,026.9)) and dual LAP (mean 4,593.6 N (SD 930.3)) demonstrated higher failure loads than diaphyseal LAP (mean 3,226.1 N (SD 423.0)) and the LCP-only control (mean 3,266.7 N (SD 676.5)) (p = 0.014 to 0.042), with no significant difference between proximal and dual LAP constructs. Composite bone distance increased progressively during cyclic loading. Fracture lines clustered near the fourth hole in the proximal LAP configuration and near the second hole in the diaphyseal and dual LAP configurations. Early fatigue failure occurred in the proximal LAP (2/6), diaphyseal LAP (1/6), and control (2/6) groups, whereas no specimen failed during cyclic loading in the dual LAP group. CONCLUSION: This biomechanical study demonstrates that proximal and dual LAP configurations provide higher static failure strength than diaphyseal LAP or LCP-only constructs, while resistance to early cyclic fracture was similar among configurations. The absence of early fatigue failure in the dual LAP configuration indicates a mechanical advantage when anatomical placement allows.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Biomechanical effects of far-cortex over-penetration in locking plate fixation: A finite element study.Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine · 2026
- Axial Compression Behaviour of Locking Plate Constructs in Different Fixation Functions in the Canine Femur.Veterinary and comparative orthopaedics and traumatology : V.C.O.T · 2026
- Locking plate fixation with cerclage augmentation and bone grafting for Vancouver B2 periprosthetic femoral fractures around cementless stems: a surgical option in elderly high-risk patients unsuitable for revision arthroplasty (mid-term outcomes).Frontiers in surgery · 2026
- A novel stiffness-distribution-controlled hybrid locking compression plate for osteoporotic midshaft clavicle fractures: a finite element study.BMC musculoskeletal disorders · 2026
- Periprosthetic femoral fractures: Ten-year clinical experience using a fracture-specific fixation system: A retrospective study.The surgeon : journal of the Royal Colleges of Surgeons of Edinburgh and Ireland · 2026
- Biomechanical comparison of prophylactic fixation methods following excision of benign aggressive tumors of the femoral neck using composite bone models.Clinical biomechanics (Bristol, Avon) · 2026