P10.10.B Optimizing Imaging Dose in Pediatric Neuro-Radiotherapy: Integrating Retrospective Analysis and Predictive Tools for IGRT Dose Reduction within the Image Gently Radiation Safety Framework
S Mariani, L Bernabucci, G Stimato, G Turco, V Pollutri, A Capotosti, E Meldolesi, F Greco,
Key points
From the paper's conclusion- Overall, implementation of the Image Gently protocol resulted in a 92.5% reduction in mean effective dose from brain IGRT imaging, with a further 1.1% reduction achieved through predictive optimisation.
Abstract
Background In paediatric oncology, improving survival rates increasingly requires careful consideration of long-term treatment-related effects, including cumulative radiation exposure. Modern radiotherapy provides highly conformal dose delivery while sparing healthy brain tissue; however, image-guided radiotherapy (IGRT) adds additional radiation dose, which is particularly relevant in paediatric patients due to their increased radiosensitivity. Material and Methods This study evaluated the feasibility and clinical impact of implementing the Image Gently protocol (Varian©) specifically for brain radiotherapy with ACS-based image guidance, supported by Python-based software tools for dose tracking and optimisation. The protocol used reduced-dose CBCT acquisition parameters (80 kV, 20 mA; CTDIvol 0.94 mGy; DLP 20.2 mGy·cm). Effective dose was estimated using ED = DLP × k (ICRP 103). A dedicated Python-based tool enabled standardised cumulative dose reporting and highlighted variability in pre-treatment imaging practices, including operator-dependent differences in acquisition settings and filter usage. In addition, a predictive software module integrating patient and treatment-related variables (age, brain treatment site, number of fractions, and acquisition mode) was developed to optimise imaging schedules. When predicted cumulative DLP exceeded a predefined diagnostic CT-equivalent threshold, the system recommended reducing CBCT frequency. Results A total of 50 paediatric patients were included. CBCT frequency was selectively reduced without compromising treatment accuracy or exceeding international imaging safety constraints. Workflow standardisation reduced operator dependence, while predictive planning ensured cumulative dose control and maintained image quality and setup precision. Overall, implementation of the Image Gently protocol resulted in a 92.5% reduction in mean effective dose from brain IGRT imaging, with a further 1.1% reduction achieved through predictive optimisation. Conclusion Integration of these tools into routine clinical workflow enables standardised, operator-independent brain IGRT management, supporting safer, more consistent, and personalised radiotherapy in paediatric neuro-oncology.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Paediatric radiotherapy in Denmark from 2008 to 2024: implementation of improved radiotherapeutic technologies.Acta oncologica (Stockholm, Sweden) · 2026
- Rethinking radiation dose optimization in pediatric computed tomography imaging: emerging algorithms and clinical integration.Pediatric radiology · 2026
- Performance of Commercial Deep Learning-Based Radiation Dose Optimization Software in Pediatric Radiology: A Systematic Review.Children (Basel, Switzerland) · 2026
- Size-specific dose estimation in pediatric computed tomography: From dose characterization to individualized optimization.World journal of radiology · 2026
- Doses in planning computed tomography in one radiotherapy department: Description of exposure parameters and estimation of organ doses, effective doses and typical doses.Radiography (London, England : 1995) · 2026
- Optimization of image quality and radiation dose in low-dose pediatric chest CT Using 80 kVp acquisition and variable ASiR-V reconstruction: A retrospective comparative study.Medicine · 2026