EXPRESS: Development of dynamic time-over-threshold method for signal processing in cavity ring-down spectroscopy.
Konrad Kleineidam, Ryohei Terabayashi, Takumi Mochizuki, Erik Thiel, Klaus Wendt, Norihiko Nishizawa, Kenji Shimazoe, Hideki Tomita
Key points
From the paper's conclusion- This work investigates a simplified signal processing approach based on the Time-over-Threshold (ToT) method, using only threshold crossing times instead of full waveform acquisition.
Abstract
Cavity ring-down spectroscopy (CRDS) enables highly sensitive detection of trace gases but typically requires high-speed, high-resolution digitizers for exponential fitting of the measured "ring-down" decay signals. This work investigates a simplified signal processing approach based on the Time-over-Threshold (ToT) method, using only threshold crossing times instead of full waveform acquisition. In contrast to the ToT method with static thresholds, dynamic threshold signals are employed to improve linearity and robustness with respect to signal amplitude variations. The proposed dynamic ToT method is implemented using a compact analog circuit and evaluated using a mid-infrared CRDS setup for CO2 measurement. Experimental results demonstrate good agreement with conventional fitting, with improved reliability compared to the ToT method with static thresholds and comparable linearity over a broad range of ring-down decay rates. These findings establish the dynamic ToT method as an accessible and scalable alternative for acquiring decay data in CRDS, particularly suited for simultaneous multi-channel or multi-wavelength measurements.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Continuous-Wave Cavity Ring-Down Spectroscopy: Instrument Architectures, Modulation Strategies, and Emerging ApplicationsSpectroscopy Journal · 2026
- Cavity Ring-Down Spectroscopy at Benchmark-Level Sub-Permille Accuracy Enabled by a System-Level Detection Transfer FunctionarXiv · 2026
- Full-range CO2 detection using dual-path WMS-DAS and multi-regime adaptive fusion.Optics letters · 2026
- Quantitative analysis of tritiated water using cavity ring-down spectroscopyScientific Reports · 2026
- Artifact-resilient design and early reading extraction in a wearable microcavity NDIR transcutaneous CO2 prototype: validation of a PDMS interface layer and the R2-max rate-window method.Biomedical engineering letters · 2026
- Towards simultaneous methane clumped-isotopologue and radiocarbon measurements by cavity-ring-down spectroscopy.Optics express · 2026