Artifact-resilient design and early reading extraction in a wearable microcavity NDIR transcutaneous CO2 prototype: validation of a PDMS interface layer and the R2-max rate-window method.
Xin Xiong, Hanzheng Xu, Yunying Cai, Xiao Mo, Heng Su, JianFeng He
Faculty of Information Engineering and Automation, Kunming University of Science and Technology, Kunming, 650500 China. · Endocrinology Department, First People's Hospital of Yunnan Province, The Affiliated Hospital of Kunming University of Science and Technology, Type 1 Diabetes Alliance of Yunnan Province, Kunming, 650032 China.
Key points
From the paper's conclusion- This study developed a wearable microcavity NDIR tcCO2 prototype and evaluated whether a PDMS interface layer combined with an R2-max rate-window method could improve signal stability and early reading extraction.
Abstract
Wearable skin-attached NDIR CO2 sensing is vulnerable to humidity, condensation, and contact perturbations, which can cause unstable signals and delayed valid readings. This study developed a wearable microcavity NDIR tcCO2 prototype and evaluated whether a PDMS interface layer combined with an R2-max rate-window method could improve signal stability and early reading extraction. The prototype integrated a miniature NDIR CO2 sensor, a 3D-printed wrist microcavity, and a wristband fixation structure. Human wrist recordings were obtained under direct detection without a membrane (NoMem) and with a PDMS membrane. Signals were preprocessed using median filtering and local polynomial smoothing. Sliding-window linear fitting was then applied, and valid rate windows were selected using R2, drop ratio, and slope constraints. Supplementary gas-gap, PDMS-thickness, repeatability, and post-exercise experiments were also performed. Valid rate windows were identified in 4/4 PDMS records and 4/8 NoMem records. The PDMS condition showed a median rise rate of 0.954 raw/s (IQR: 0.105 raw/s) and a median window center time of 280.5 s, whereas NoMem showed 1.159 raw/s (IQR: 0.581 raw/s) and 360.0 s. PDMS yielded more concentrated rate distributions and fewer unstable or invalid outcomes. Supplementary analyses supported the 0.3 cm gas gap, 100 μm PDMS film, repeated valid-window extraction, and post-exercise early-window formation. The PDMS interface and R2-max method improved prototype-level raw-signal stability and interpretable early-window extraction. These findings support engineering feasibility for wearable microcavity NDIR CO2 monitoring, but do not yet establish clinical quantitative tcCO2 accuracy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Humidity-Strain Synergized Adaptive NO Sensing in a Pd-Carbon Nanocoil Fabric for Respiratory-State Recognition.ACS applied materials & interfaces · 2026
- A Fully Integrated Smart Facemask for Ultrafast and Humidity-Resistant Exhaled CO2 Monitoring.ACS sensors · 2026
- EXPRESS: Development of dynamic time-over-threshold method for signal processing in cavity ring-down spectroscopy.Applied spectroscopy · 2026
- Polymer Optofluidic Chip for Double-Beam Photometric Determination of Nitrite in Water Samples.ACS omega · 2026
- Near-Field Communication-Enabled Wearable Chemical Sensors for Bioanalytical Monitoring.Chemistry, an Asian journal · 2026
- An All-Polyurethane Nanomesh Triboelectric Sensor for Swallowing and Voice Detection.ACS nano · 2026