Towards simultaneous methane clumped-isotopologue and radiocarbon measurements by cavity-ring-down spectroscopy.
Mehr Fatima, Ville Ulvila, Ivan Prokhorov, Gang Li, Joachim Mohn, Thomas Hausmaninger
Key points
From the paper's conclusion- Our results demonstrate that CRDS provides a sensitive and reliable approach for monitoring doubly substituted methane isotopologues on a shared platform with the capability for radiocarbon detection.
Abstract
Methane (CH4) is a potent greenhouse gas, and proper source attribution is essential for effective climate change mitigation. Traditional measurement techniques, such as singly substituted isotope ratio measurements (δ13C and δD) by IRMS or optical spectroscopy, often struggle to distinguish between different methane sources because their isotopic ratios can overlap. Doubly substituted isotope analysis, commonly referred to as clumped isotope analysis, offers improved differentiation. However, these isotopologues are challenging to detect due to their low abundance. In this study, we demonstrate the detection of a doubly substituted methane isotopologue (13CH3D) using cavity-ring-down spectroscopy (CRDS) at a wavelength of 4.5 µm, a region also used for the detection of 14CO2. We further present a comprehensive data analysis routine for our CRDS measurements, including higher order mode filtering based on signal selection criteria, and relative and absolute frequency calibration. The approach is general and can be applied to other CRDS systems. Spectral lines recently assigned to 13CH3D were addressed and characterized. The system achieves a precision of 19‰ at an averaging time of 50 minutes for 13CH3D. Our results demonstrate that CRDS provides a sensitive and reliable approach for monitoring doubly substituted methane isotopologues on a shared platform with the capability for radiocarbon detection.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Simultaneous High-Precision¹³ C Isotope Ratio Analysis of CO₂ and CH₄ Using an FDM-OA-CEAS SensorAnalytical Chemistry · 2026
- Simultaneous Detection of 12CH4, 13CH4, and Related Isotope Ratio Exploiting a Frequency-Multiplexed Mid-Infrared Quartz-Enhanced Photoacoustic Sensor.ACS sensors · 2026
- Line profile study and intensity depletion of the R-branch manifolds in the 2ν3 band of 12CH4 in dry airJournal of Quantitative Spectroscopy and Radiative Transfer · 2026
- A reconfigurable and compact circular multi-pass sensor for ultra-sensitive methane detectionJournal of Instrumentation · 2026
- Long‐Term Performance and Bulk‐Isotope Effects in Methane Clumped‐Isotope Measurements Using a 253 Ultra HR‐IRMSRapid Communications in Mass Spectrometry · 2026
- LITES-based methane isotope ratio measurement and source attributionOptics Express · 2025