Influence of Room Illumination on Corneal Densitometry.
Miriam Esteban-Burgués, Ana R Arizcuren, Cristina Pelayo, Alejandra Consejo
Aragon Institute for Engineering Research, Universidad de Zaragoza, Zaragoza, Spain.
Key points
From the paper's conclusion- Scheimpflug-based corneal densitometry is highly sensitive to room illumination, even within clinically plausible lighting conditions, whereas the OCT-derived metrics are robust to such variations.
Abstract
PURPOSE: To evaluate the effect of room illumination on corneal densitometry measurements obtained with Scheimpflug imaging and optical coherence tomography (OCT) and on corneal geometric parameters. METHODS: Twenty-five healthy participants (50 eyes) were imaged under four controlled illumination conditions: extreme scotopic (0.003 lx), dim-light/mesopic (30 lx), low photopic (130 lx), and moderate photopic (220 lx) conditions. Corneal densitometry was quantified as the mean pixel intensity for both the Scheimpflug and the OCT images, with speckle contrast additionally computed for the OCT images. Corneal geometry (central corneal thickness and keratometry readings defined as corneal power in the flattest [K1] and steepest [K2] meridians) was extracted from Scheimpflug tomography. Linear mixed-effects models were used to assess the effect of illumination on corneal parameters. RESULTS: Scheimpflug-derived corneal densitometry increased significantly with illumination (P < 0.001), rising from 45.8 ± 4.3 arbitrary units under extreme scotopic conditions to 69.5 ± 11.5 arbitrary units under moderate photopic illumination. Significant differences were observed between all illumination levels (all P < 0.01). In contrast, the OCT-derived mean pixel intensity and speckle contrast remained stable across conditions (P > 0.05). The corneal geometric parameters showed no significant changes with room illumination (all P > 0.05). CONCLUSIONS: Scheimpflug-based corneal densitometry is highly sensitive to room illumination, even within clinically plausible lighting conditions, whereas the OCT-derived metrics are robust to such variations. TRANSLATIONAL RELEVANCE: Room illumination-induced changes may bias corneal densitometry readings and should be carefully controlled and reported in densitometry studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantitative assessment of corneal haze: comparative optical densitometry using anterior segment OCT and Pentacam.International ophthalmology · 2026
- Corneal densitometry and anterior segment changes in patients undergoing maintenance hemodialysis.International ophthalmology · 2026
- Total Corneal Aberration Measurements by Anterior Segment Optical Coherence Tomography and Scheimpflug-Based Imaging: A Narrative Review of Optical Principles and Sources of Inter-Device Discordance.Journal of clinical medicine · 2026
- Tear Dynamics During Fenestrated Scleral Lens Wear: A Pilot StudyOphthalmic and Physiological Optics · 2026
- Agreement of corneal thickness measured by Scheimpflug, SS-OCT, and specular microscopy in Chinese myopia.BMC ophthalmology · 2026
- Determinants of posterior corneal optical densitometry and their relationship with endothelial morphology.Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie · 2026