Evaluation of Neuroangiography Settings and Protective Shielding on Radiation Exposure to Patient and Interventionist: A Comparative Dosimetric Study.
Zahra Azizan, Neda Izadi, Binesh Ninan, Mohsen Nouri, Pouria Moshayedi
Iranian Center of Neurological Research, Neuroscience Institute, Imam Khomeini Hospital Complex (Z.A., P.M.), Tehran University of Medical Sciences, Iran. · Research Center for Social Determinants of Health, Research Institute for Metabolic and Obesity Disorders, Research Institute for Endocrine Sciences, Shahid Beheshti University of Medical Sciences, Tehran, Iran (N.I.).
Key points
From the paper's conclusion- Low frame rates, close image intensifier positioning, minimal magnification, and double shielding substantially reduce radiation risk for both operators and patients. These evidence-based configurations provide a practical framework for safer neuroendovascular procedures.
Abstract
BACKGROUND: Neuroendovascular procedures expose surgeons and patients to significant radiation, increasing long-term health risks. Despite the known effects of imaging parameters on radiation exposure, the relative impacts of different imaging and shielding strategies have been less studied. We aimed to assess the effects of digital subtraction angiography parameter modifications on operator and patient radiation doses. METHODS: Using a phantom skull in a neuroendovascular suite, radiation was measured under varying digital subtraction angiography settings, including image intensifier distance, collimation, magnification, frame rate, and protective shielding. Operator exposure was recorded using a dosimeter, and patient dose was obtained from the angiography system. Each configuration was tested in 3 independent trials. One-way ANOVA was used to determine statistical significance. RESULTS: Lowering the frame rate (5.2-fold) and moving the image intensifier closer to the patient (3.5-fold) significantly reduced operator exposure (P<0.0001). Double shielding reduced operator dose 10-fold (P<0.0001). Patient dose was minimized by reducing magnification (9.3-fold; P<0.0001), lowering the frame rate (2.4-fold; P<0.0001), and decreasing image intensifier distance (1.7-fold; P<0.0001). Collimation decreased operator dose but slightly increased patient skin dose. Optimal parameter combinations maintained operator exposure well below recommended safety limits while minimizing patient dose. CONCLUSIONS: Low frame rates, close image intensifier positioning, minimal magnification, and double shielding substantially reduce radiation risk for both operators and patients. These evidence-based configurations provide a practical framework for safer neuroendovascular procedures.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Impact of examination geometry on radiation dose to patient and operator during interventional procedures.Irish journal of medical science · 2026
- Impact of Angiography System Generation on Radiation Exposure in Peripheral, Abdominal, and Cerebral Angiography: Comparison Across Three Consecutive Generations.RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin · 2026
- Mobile radiation shielding device for scatter dose reduction during portable radiography in an orthopedic ward.PloS one · 2026
- Radiation Dose Metrics and Local Diagnostic Reference Levels in Low-Dose Stent-Assisted Coiling of Intracranial AneurysmsJournal of Clinical Medicine · 2026
- Evaluation of radiation-protection drapes for reducing operator and patient radiation dose during overtable fluoroscopy: a phantom study.Journal of radiological protection : official journal of the Society for Radiological Protection · 2026
- Clinical evaluation of radiation-shielding effects of a head-and-neck X-ray protective device during cardiovascular interventional procedures.Radiological physics and technology · 2026