Radiation Dose Metrics and Local Diagnostic Reference Levels in Low-Dose Stent-Assisted Coiling of Intracranial Aneurysms
Mariusz Stanisław Sowa, Joanna Sowa, Kamil Adam Węglarz, M. Budzanowski
University of Warmia and Mazury in Olsztyn · University of Computer Sciences and Economics in Olsztyn
Key points
From the paper's conclusion- Background/Objectives: Operator experience, the implementation of low frame rates during both fluoroscopy and digital subtraction angiography (DSA), and the use of modern angiographic systems are essential for maintaining diagnostic image quality while minimizing ionizing radiation exposure during stent-assisted endovascular treatment of intracranial aneurysms.
Abstract
Background/Objectives: Operator experience, the implementation of low frame rates during both fluoroscopy and digital subtraction angiography (DSA), and the use of modern angiographic systems are essential for maintaining diagnostic image quality while minimizing ionizing radiation exposure during stent-assisted endovascular treatment of intracranial aneurysms. At the study center, a low-dose protocol is employed, using the lowest available fluoroscopy frame rate (3.125 frames per second) and a nominal acquisition rate of 2 frames per second for DSA, three-dimensional (3D) rotational angiography, 2D/3D mapping, and roadmapping. Methods: A retrospective analysis was performed on 132 stent-assisted procedures conducted at a single tertiary center between 2018 and 2024. For each procedure, data were collected for dose-area product (DAP), reference air kerma (Ka,r), fluoroscopy time (FT), and the total number of DSA frames. Local diagnostic reference levels (DRLs; 75th percentile [P75]) and typical values (50th percentile [P50]) were established and compared with values reported in the literature. Results: For all patients the P75 values, representing DRLs, were 19.89 Gy·cm2 for DAP, 332 mGy for Ka,r, 25 min 32 s for FT, and 354 DSA frames. The P50 values were 13.71 Gy·cm2 for DAP, 219.5 mGy for Ka,r, 20 min 36 s for FT, and 277 DSA frames. Conclusions: In this single-center cohort, dose metrics for stent-assisted coil embolization were within the lower range of published values. Cross-study comparisons remain descriptive and require cautious interpretation. The proposed local DRLs may support quality assurance, dose optimization, and patient safety in similar clinical settings. Further multicenter and multi-operator studies are necessary to assess transferability and applicability beyond coil-only procedures. Limitations include the retrospective single-center design (single operator) and the lack of a contemporaneous control group and formal image-quality/outcome assessment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Patient radiation dose in diagnostic cerebral angiography for intracranial aneurysm: local diagnostic reference levels from a single-center study in Thailand.Radiation protection dosimetry · 2026
- Local Diagnostic Reference Levels and Objective Image Quality in Mechanical Thrombectomy Performed with a Low-Dose Protocol: A Single-Center Observational Study.Journal of clinical medicine · 2026
- Radiation Exposure in Acute Endovascular Stroke Treatment: A Focus on Mechanical ThrombectomyNeuroimaging · 2026
- Real-time dosimetry in interventional radiology—comparing the operators’ radiation dose and associated factors for five types of transarterial embolization proceduresFrontiers in Public Health · 2026
- Efficiency and Optimization of Iodinated Contrast Agents in Cerebral Aneurysm CT Angiography: A Systematic Review.Diagnostics (Basel, Switzerland) · 2026
- Operator Radiation Dose in C-Arm Fluoroscopy: A Narrative Review of Region- and Procedure-Specific Evidence and a Practical Framework for Intraoperative Dose EstimationSurgeries · 2026