Impact of examination geometry on radiation dose to patient and operator during interventional procedures.
Lucie Sukupova
Institute of Clinical and Experimental Medicine, Videnska 1958/9, Prague, Czech Republic. Lucie.sukupova@ikem.cz.
Key points
From the paper's conclusion- Examination geometry influences both patient dose indicators and operator dose. Reducing X-ray field size decreases PKA and operator dose, although air kerma at the IRP may increase slightly. Within the geometries investigated, a shorter patient-to-detector distance was associated with lower operator dose. Identical PKA values do not necessarily imply identical operator doses.
Abstract
BACKGROUND: In interventional radiology, staff radiation exposure mainly originates from patient-scattered radiation. While standard protective tools are used, examination geometry, especially detector positioning and X-ray field size, critically influences both patient and operator doses. Quantitative data on these effects under controlled experimental conditions are limited. AIMS: To quantify the influence of examination geometry on patient and operator radiation doses and identify configurations minimising radiation exposure. METHODS: Measurements were performed using an Artis Q angiography system and polymethyl methacrylate phantoms measuring 17 cm and 25 cm in thickness, positioned at a height of 95 cm above the floor to simulate abdominal geometry. Kerma-area product (PKA), air kerma at the interventional reference point (IRP), and operator personal dose equivalent Hp(10) measured at heights of 105, 135, and 165 cm above the floor, were recorded for different phantom-to-detector distances and X-ray field sizes. Operator dose was measured using the RaySafe i3 system. RESULTS: Increasing phantom thickness increased operator dose by an average factor of 4.0 (range, 1.8-6.8). For the large phantom and large X-ray field, air kerma at the IRP increased from 9.8 to 12.9 mGy per acquisition as the phantom-to-detector distance increased. Operator dose increased by up to 77% with a larger X-ray field and by more than 60% with increasing phantom-to-detector distance. Identical PKA values were obtained for different examination geometries, despite markedly different operator doses. Dose readings measured on top of the detector were below the detection limit of the dosimetry system in all measurements. CONCLUSION: Examination geometry influences both patient dose indicators and operator dose. Reducing X-ray field size decreases PKA and operator dose, although air kerma at the IRP may increase slightly. Within the geometries investigated, a shorter patient-to-detector distance was associated with lower operator dose. Identical PKA values do not necessarily imply identical operator doses.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of Neuroangiography Settings and Protective Shielding on Radiation Exposure to Patient and Interventionist: A Comparative Dosimetric Study.Stroke (Hoboken, N.J.) · 2026
- Mobile radiation shielding device for scatter dose reduction during portable radiography in an orthopedic ward.PloS one · 2026
- Evaluation of radiation-protection drapes for reducing operator and patient radiation dose during overtable fluoroscopy: a phantom study.Journal of radiological protection : official journal of the Society for Radiological Protection · 2026
- Real-time dosimetry in interventional radiology—comparing the operators’ radiation dose and associated factors for five types of transarterial embolization proceduresFrontiers in Public Health · 2026
- Influence of collimation on radiation exposure during chest radiography.Radiation protection dosimetry · 2026
- Operator Radiation Dose in C-Arm Fluoroscopy: A Narrative Review of Region- and Procedure-Specific Evidence and a Practical Framework for Intraoperative Dose EstimationSurgeries · 2026