Ultraviolet-C Light Emitter as a Method for Reducing Bacterial Contamination of Ophthalmic Instruments.
Raul Velez-Montoya, K Carolina Franco-Ramirez, Abril T Vargas-Fernández, Jesus E Arreola-Martinez, J Armando Cruz-Cervantes, Virginia Vanzzini-Zago, Jans Fromow-Guerra
Retina Department, Asociación para Evitar la Ceguera en México IAP, Mexico City 04030, Mexico, apec.org.mx. · Microbiology Department, Asociación para Evitar la Ceguera en México IAP, Mexico City 04030, Mexico, apec.org.mx.
Key points
From the paper's conclusion- This exploratory laboratory study suggests that, after 5 min of UVC exposure using a commercially available emitter, no recoverable bacteria were detected on lid speculums under these experimental conditions. Although spores, fungi, and viruses were not evaluated, these findings suggest that UVC irradiation may be a feasible approach for reducing bacterial contamination of reusable instruments during in-office ophthalmic procedures.
Abstract
BACKGROUND: To assess the efficacy of an ultraviolet light type C (UVC) emitter in reducing bacterial load in common surgical equipment used during in-office ophthalmological procedures. METHODS: We used 7 steam-sterilized lid speculums. The first one was the negative control. The second was left unprotected in a high-traffic area for 24 h before being cultivated. Lid speculums 3 to 7 were contaminated with a strain of Staphylococcus epidermidis, recovered from an active case of postoperative endophthalmitis. The third one was cultivated immediately after contamination. Lid speculums 4 to 7 were exposed to UVC for 30 s, 2, 5, and 10 min, respectively (continuous 254 nm UVC, irradiance > 160 mW/cm2). All speculums were cultivated on Mueller-Hinton agar plates for 48 h (≈21% O2). Differences in bacterial load, observed colonies, and colony-forming units were assessed with a Kruskal-Wallis test. All experiments were done in triplicate. RESULTS: We analyzed 27 bacterial cultures. After 24 h, lid speculum 2 showed growth of a single colony of Sphingomonas paucimobilis. After 30 s of UVC exposure, bacterial growth was 108 ± 7.2 CFU (bacterial load: 3240 ± 214.7). Bacterial growth decreased progressively with increasing UVC exposure, reaching no detectable growth after 5 min, representing a 100% reduction (p < 0.01). A second culture obtained from lid speculum 2 and the positive control after additional 5 min of UVC exposure also showed no bacterial growth. CONCLUSIONS: This exploratory laboratory study suggests that, after 5 min of UVC exposure using a commercially available emitter, no recoverable bacteria were detected on lid speculums under these experimental conditions. Although spores, fungi, and viruses were not evaluated, these findings suggest that UVC irradiation may be a feasible approach for reducing bacterial contamination of reusable instruments during in-office ophthalmic procedures.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- UV-C LED Disinfection of Flexible Rhinolaryngoscopes: Microbiological, Economic, Environmental, and Legislative Evaluation.Laryngoscope investigative otolaryngology · 2026
- Inactivation of microorganisms in the complex regions of transvaginal ultrasound probes by a UVC-LED light based disinfection system.PloS one · 2026
- Environmental Impact of a Novel UV-C Disinfection Machine for Non-Channel Flexible Laryngoscopes.Laryngoscope investigative otolaryngology · 2026
- Simulation-Based Implementation of Germicidal UV-C Light Within a GMP SettingHygiene · 2026
- UV-C disinfection of endocavitary ultrasound probes against Candidozyma auris : a rapid and chemical-free solutionAntimicrobial Stewardship & Healthcare Epidemiology · 2026
- Radiometrically Equivalent Comparison of 254 nm Low-Pressure Mercury Lamps and InGaN 280 nm UV-C LEDs for Surface Disinfection.ACS omega · 2026