UV-C LED Disinfection of Flexible Rhinolaryngoscopes: Microbiological, Economic, Environmental, and Legislative Evaluation.
Camille Maeyaert, Sofie Claeys, Stien Vandendriessche, Hannah Siwe, Bjorn Delbeecke, Pascal Verdonck, Alain F Kalmar
Faculty of Medicine and Health Sciences Ghent University Ghent Belgium. · Department of Otorhinolaryngology Ghent University Hospital Ghent Belgium.
Key points
From the paper's conclusion- The Ray-Two UV-C LED chamber achieves near-complete bacterial elimination (100% in 19/20 samples; 1 residual CFU in 1/20 samples) on flexible rhinolaryngoscopes, with superior cost-efficiency and sustainability versus TTWS. This mercury-free, point-of-care method supports greener ORL practice amid evolving EU and US state-level regulations on mercury-containing lamps, enhancing infection control without compromising efficacy.
Abstract
OBJECTIVE: To evaluate the microbiological efficacy of the Ray-Two ultraviolet-C (UV-C) light-emitting diode (LED) chamber for reprocessing flexible rhinolaryngoscopes in otorhinolaryngology (ORL) practice and compare its economic and ecological impacts with chemical high-level disinfection using the Tristel Trio Wipes System (TTWS). METHODS: In a prospective study at a tertiary ORL center, bacterial contamination was assessed on 20 flexible rhinolaryngoscopes immediately after routine posterior rhinoscopy and/or laryngoscopy. Paired swab samples were collected from the distal 10-12 cm shaft pre- and post-reprocessing: manual pre-cleaning with tap water followed by a 4-min UV-C LED cycle. Samples were cultured on TSA-5% sheep blood agar for 36 h at 35°C-37°C, with colony-forming units (CFUs) enumerated. Per-cycle costs and carbon footprints were calculated via life cycle assessment (ISO 14040/14044) and compared with TTWS, based on 5000 annual procedures. RESULTS: Pre-reprocessing median CFUs (IQR; range) were 15 (3-27; 2 to > 200), reducing to 0 (0-0; 0-1) post-UV-C (p < 0.001; > 99% reduction; 19/20 samples sterile). UV-C per-cycle cost was €0.56 (vs. €8.32 for TTWS; 93% savings), with a carbon footprint of 6.81 g CO2e (vs. 115.8 g; 94% reduction). Annual savings: €38,800 and 548 kg CO2e. CONCLUSION: The Ray-Two UV-C LED chamber achieves near-complete bacterial elimination (100% in 19/20 samples; 1 residual CFU in 1/20 samples) on flexible rhinolaryngoscopes, with superior cost-efficiency and sustainability versus TTWS. This mercury-free, point-of-care method supports greener ORL practice amid evolving EU and US state-level regulations on mercury-containing lamps, enhancing infection control without compromising efficacy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Environmental Impact of a Novel UV-C Disinfection Machine for Non-Channel Flexible Laryngoscopes.Laryngoscope investigative otolaryngology · 2026
- Ultraviolet-C Light Emitter as a Method for Reducing Bacterial Contamination of Ophthalmic Instruments.Journal of ophthalmology · 2026
- Effectiveness of ultraviolet-C robots in isolation room disinfection: Impact of dose, distance, and direct exposure on environmental surface decontamination.PloS one · 2026
- Inactivation of microorganisms in the complex regions of transvaginal ultrasound probes by a UVC-LED light based disinfection system.PloS one · 2026
- Germicidal Ultraviolet C (UV-C) Light for Surface Disinfection in Hospitals: Mapping the Evidence on Devices, Parameters, Effectiveness, and ImplementationHygiene · 2026
- UV-C disinfection of endocavitary ultrasound probes against Candidozyma auris : a rapid and chemical-free solutionAntimicrobial Stewardship & Healthcare Epidemiology · 2026