Environmental Impact of a Novel UV-C Disinfection Machine for Non-Channel Flexible Laryngoscopes.
Abhishek Mahesh, Paul C Bryson
Case Western Reserve University School of Medicine Cleveland Ohio USA. · Department of Otolaryngology-Head and Neck Surgery Cleveland Clinic Lerner College of Medicine Cleveland Ohio USA.
Key points
From the paper's conclusion- Environmental analysis demonstrates a significant amount of usage of disposables, water, and energy per non-channeled nasolaryngoscope with current HLD protocols. UV-C HLD resulted in less consumption of these resources. The marked reduction in turnover time and scope handling represents a significant opportunity for clinical efficiency, cost reduction, and resource consumption and will merit further investigation when FDA approval is granted.
Abstract
OBJECTIVES: Infection prevention in hospitals is a highly discussed and regulated process for endoscopy in Otolaryngology. The current standard for endoscopes involves high level disinfection (HLD). The UV-SMART D60 disinfection machine utilizes Impelux UV-C light technology. This study aimed to evaluate the environmental impact of the D60 machine within an otolaryngology department in comparison to current HLD protocols. METHODS: Three equipment, materials, and disposables audits were conducted where the turnover and HLD process of a single non-channeled laryngoscope was analyzed. Every item involved in the cleaning process was documented for both standard HLD and D60 workflows. Water quantity was documented and a power meter was used to determine the wattage of the machinery used. Additionally, scope turnover time was documented and CO2 emissions were calculated from known quantities. RESULTS: Results demonstrated a significant reduction in material usage with the D60 including reduced PPE. The total water usage was about 45 L more per scope for traditional HLD. Power analysis demonstrated a reduction of 250 watts per scope. Additionally, the D60 reduced CO2 emissions by over a factor of 3500. Scope turnover time was 109 min less for the D60. CONCLUSIONS: Environmental analysis demonstrates a significant amount of usage of disposables, water, and energy per non-channeled nasolaryngoscope with current HLD protocols. UV-C HLD resulted in less consumption of these resources. The marked reduction in turnover time and scope handling represents a significant opportunity for clinical efficiency, cost reduction, and resource consumption and will merit further investigation when FDA approval is granted. LEVEL OF EVIDENCE: N/A.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- UV-C LED Disinfection of Flexible Rhinolaryngoscopes: Microbiological, Economic, Environmental, and Legislative Evaluation.Laryngoscope investigative otolaryngology · 2026
- UV-C disinfection of endocavitary ultrasound probes against Candidozyma auris : a rapid and chemical-free solutionAntimicrobial Stewardship & Healthcare Epidemiology · 2026
- Germicidal Ultraviolet C (UV-C) Light for Surface Disinfection in Hospitals: Mapping the Evidence on Devices, Parameters, Effectiveness, and ImplementationHygiene · 2026
- Ultraviolet-C Light Emitter as a Method for Reducing Bacterial Contamination of Ophthalmic Instruments.Journal of ophthalmology · 2026
- Effectiveness of ultraviolet-C robots in isolation room disinfection: Impact of dose, distance, and direct exposure on environmental surface decontamination.PloS one · 2026
- Radiometrically Equivalent Comparison of 254 nm Low-Pressure Mercury Lamps and InGaN 280 nm UV-C LEDs for Surface Disinfection.ACS omega · 2026