Copper filtration reduces radiation dose while maintaining image quality in cerebral flat-detector CT.
Niclas Schmitt, Tim Hilgenfeld, Dominik F Vollherbst, Christoph M Mooshage, Fabian Preisner, Michael O Breckwoldt, Alexander Hubert, Daniel Schwarz,
Neuroradiology, University Hospital Heidelberg, Heidelberg, Germany niclas.schmitt@med.uni-heidelberg.de. · Neuroradiology, University Hospital Heidelberg, Heidelberg, Germany.
Key points
From the paper's conclusion- Cu filtration in cerebral FDCT enables substantial radiation dose reduction while preserving image quality and intracranial hemorrhage detection, supporting its clinical implementation as a practical dose optimization strategy for peri-interventional imaging.
Abstract
BACKGROUND: Flat-detector computed tomography (FDCT) is increasingly used for peri-interventional cerebral imaging but is associated with a relatively high radiation exposure. Copper (Cu) filtration may reduce radiation dose. However, its impact on cerebral image quality and intracranial hemorrhage detection remains unclear. METHODS: In this retrospective single-center study, 31 patients undergoing neurointerventional procedures with intraindividual FDCT acquisitions with and without Cu filtration were analyzed. Quantitative image quality was assessed using contrast-to-noise ratio (CNR). Qualitative image analysis and intracranial hemorrhage detection were independently evaluated by two readers blinded to Cu filtration status using five-point scales. RESULTS: Cu filtration resulted in a significant radiation dose reduction of 25.9% for both entrance skin dose (145.19±13.18 mGy vs 195.89±18.05 mGy) and dose-area product (41.52±3.77 Gy·cm² vs 56.01±5.16 Gy·cm²), respectively (P<0.001). No differences in CNR were observed for unfiltered vs Cu-filtered FDCT (basal ganglia: 4.73±2.04 vs 4.37±1.99, P=0.419). Qualitative image ratings were similar between techniques (supratentorial cortex: 2.27±0.66 vs 2.08±0.75, P=0.089), with very good inter-reader agreement (κ=0.86; 95% CI: 0.80 to 0.91). All intracranial hemorrhages were correctly identified by both techniques. Correct exclusion of intracranial hemorrhage was 15/16 with Cu filtration and 14/16 without Cu filtration, without statistically significant difference. Differences were limited to hemorrhage mimics (n=2) and minor variations in diagnostic confidence without affecting binary classification. CONCLUSION: Cu filtration in cerebral FDCT enables substantial radiation dose reduction while preserving image quality and intracranial hemorrhage detection, supporting its clinical implementation as a practical dose optimization strategy for peri-interventional imaging.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Optimization of reconstruction strategies for head and neck computed tomography angiography using photon-counting computed tomography: A comparative study with energy-integrating detector computed tomography.European journal of radiology · 2026
- Transition from a detector-dose-based to a CNR-based cerebral angiography system: evaluation of image quality and radiation dose.Radiological physics and technology · 2026
- Photon-counting detector CT for CT pulmonary angiography: a systematic review and meta-analysis.European radiology · 2026
- Photon-Counting CT for Postoperative Imaging of Cochlear Implants: Comparison to Flat-Detector CT.Investigative radiology · 2026
- Resolution-noise characteristics of common FDK filter kernels: A practical reference for preclinical cone-beam micro-CT.PloS one · 2026
- K‐Edge imaging using a clinical dual‐source photon‐counting CT systemJournal of Applied Clinical Medical Physics · 2026