Resolution-noise characteristics of common FDK filter kernels: A practical reference for preclinical cone-beam micro-CT.
Falk L Wiegmann, Nancy L Ford
Department of Physics and Astronomy, The University of British Columbia, Vancouver, British Columbia, Canada.
Key points
From the paper's conclusion- We reconstruct identical data from a GE eXplore CT 120 scanner using four filter kernels (ramp, Shepp-Logan, cosine, Hamming) at four cutoff frequencies (1.0, 0.8, 0.6, and 0.379× Nyquist, matched to the detector-to-voxel size ratio) and evaluate each of the sixteen configurations using the modulation transfer function (MTF), noise power spectrum (NPS), and non-prewhitening detectability index (NP
Abstract
The ramp filter kernel and cutoff frequency are fundamental parameters of the Feldkamp-Davis-Kress (FDK) algorithm that determine the resolution and noise characteristics of the reconstructed image. Despite their importance, the filter configuration is frequently unreported in preclinical micro-CT studies. We reconstruct identical data from a GE eXplore CT 120 scanner using four filter kernels (ramp, Shepp-Logan, cosine, Hamming) at four cutoff frequencies (1.0, 0.8, 0.6, and 0.379× Nyquist, matched to the detector-to-voxel size ratio) and evaluate each of the sixteen configurations using the modulation transfer function (MTF), noise power spectrum (NPS), and non-prewhitening detectability index (NPW d'). Qualitative assessment is performed on a mouse lung specimen. Across the sixteen configurations, MTF10 ranges from 1.74 to 3.22 lp/mm, integrated NPS from 75 670 to 13 259 HU2, and the Rose criterion crossing diameter from 2.03 to 0.90 mm at ΔC=500 HU and from 6.89 to 3.56 mm at 100 HU. This note presents the data as a concise, primarily comparative visual and quantitative reference for FDK filter selection in preclinical cone-beam CT.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Task-based detectability and angular-sampling bounds on spatial resolution in cone-beam micro-CT: a closed-form scan-design analysis.Medical physics · 2026
- Comparison of spectral performance of three dual-energy CT scanners equipped with a deep-learning image reconstruction algorithm and one photon counting CT scanner: A phantom studyDiagnostic and Interventional Imaging · 2025
- Optimization of the reconstruction kernel for temporal bone imaging using photon-counting detector CT: A combined physical and visual evaluationRadiography · 2025
- An experimental method for direct measurement of CT detector presampling MTF from reconstructed imagesMedical Physics · 2026
- Assessment of image quality and detectability in computed tomography: Inter-laboratory comparison of Fourier-based metrics.Journal of applied clinical medical physics · 2026
- Air-Scan-Based Point-Spread-Function Estimation for Megavoltage CT with Flat-Panel DetectorsarXiv · 2026