Transition from a detector-dose-based to a CNR-based cerebral angiography system: evaluation of image quality and radiation dose.

Kota Hachiya, Hiroki Abiko, Koichi Chida, Yoshikazu Matsuda

Department of Radiology, Yamagata City Hospital Saiseikan, 1-3-26, Nanukamachi, Yamagata, Yamagata, 990-2481, Japan. hachiya.saiseikan@gmail.com. · Department of Radiology, Yamagata City Hospital Saiseikan, 1-3-26, Nanukamachi, Yamagata, Yamagata, 990-2481, Japan.

Key points

From the paper's conclusion
  • Compared with the conventional system, the new angiography platform demonstrated superior image quality under phantom conditions and was associated with a 20% reduction in total Ka,r during cerebral aneurysm EVT.

Abstract

To evaluate the image quality and clinical radiation dose of a cerebral angiography system featuring contrast-to-noise ratio (CNR)-based exposure control for neurovascular endovascular treatment (EVT). Image quality was evaluated using a quality control phantom under fluoroscopy and digital subtraction angiography (DSA). Clinical radiation dose was retrospectively compared between a conventional angiography system using detector dose-based exposure control (61 cerebral aneurysm EVT procedures) and a new angiography system using CNR-based exposure control (99 procedures). Patient entrance reference air kerma (Ka,r) and air kerma-area product (PKA) were analyzed. The new system demonstrated significantly higher observer scores than the conventional system for both fluoroscopy and DSA, including spatial resolution, low-contrast resolution, and dynamic range. The median total Ka,r was 1818 [1301-2744] mGy with the new system and 2283 [1580-3184] mGy with the conventional system, representing a 20% reduction (p < 0.05). Conversely, the median total PKA increased by 26% (173 [120-237] vs. 137 [93-207] Gy·cm2, p < 0.05), but remained below the current Japanese diagnostic reference level for cerebral aneurysm EVT. Compared with the conventional system, the new angiography platform demonstrated superior image quality under phantom conditions and was associated with a 20% reduction in total Ka,r during cerebral aneurysm EVT. Although the total PKA increased by 26%, this increase may have reflected differences in the magnification strategy between the two systems, suggesting a need for further optimization.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers