Transition from a detector-dose-based to a CNR-based cerebral angiography system: evaluation of image quality and radiation dose.
Kota Hachiya, Hiroki Abiko, Koichi Chida, Yoshikazu Matsuda
Department of Radiology, Yamagata City Hospital Saiseikan, 1-3-26, Nanukamachi, Yamagata, Yamagata, 990-2481, Japan. hachiya.saiseikan@gmail.com. · Department of Radiology, Yamagata City Hospital Saiseikan, 1-3-26, Nanukamachi, Yamagata, Yamagata, 990-2481, Japan.
Key points
From the paper's conclusion- Compared with the conventional system, the new angiography platform demonstrated superior image quality under phantom conditions and was associated with a 20% reduction in total Ka,r during cerebral aneurysm EVT.
Abstract
To evaluate the image quality and clinical radiation dose of a cerebral angiography system featuring contrast-to-noise ratio (CNR)-based exposure control for neurovascular endovascular treatment (EVT). Image quality was evaluated using a quality control phantom under fluoroscopy and digital subtraction angiography (DSA). Clinical radiation dose was retrospectively compared between a conventional angiography system using detector dose-based exposure control (61 cerebral aneurysm EVT procedures) and a new angiography system using CNR-based exposure control (99 procedures). Patient entrance reference air kerma (Ka,r) and air kerma-area product (PKA) were analyzed. The new system demonstrated significantly higher observer scores than the conventional system for both fluoroscopy and DSA, including spatial resolution, low-contrast resolution, and dynamic range. The median total Ka,r was 1818 [1301-2744] mGy with the new system and 2283 [1580-3184] mGy with the conventional system, representing a 20% reduction (p < 0.05). Conversely, the median total PKA increased by 26% (173 [120-237] vs. 137 [93-207] Gy·cm2, p < 0.05), but remained below the current Japanese diagnostic reference level for cerebral aneurysm EVT. Compared with the conventional system, the new angiography platform demonstrated superior image quality under phantom conditions and was associated with a 20% reduction in total Ka,r during cerebral aneurysm EVT. Although the total PKA increased by 26%, this increase may have reflected differences in the magnification strategy between the two systems, suggesting a need for further optimization.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Prospective study to evaluate radiation dose and image quality in two-dimensional (2D) lower limb angiographic images acquired with a novel contrast-to-noise ratio (CNR)-driven exposure control protocol.Clinical radiology · 2026
- Impact of Angiography System Generation on Radiation Exposure in Peripheral, Abdominal, and Cerebral Angiography: Comparison Across Three Consecutive Generations.RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin · 2026
- Ultra-Low-Dose Noise-Free Technology to Reduce Radiation Exposure During Coronary AngiographyCirculation Cardiovascular Interventions · 2026
- A wide-detector CT protocol for one-stop cardio-cerebrovascular angiography: a technical feasibility study comparing an integrated protocol with conventional two-scan approach.Frontiers in radiology · 2026
- Exploring a Novel Processing Algorithm to Improve Image Quality and Reduce X-Ray and Contrast Dose for Coronary Angiography: A Blinded Pilot Study.Journal of the Society for Cardiovascular Angiography & Interventions · 2026
- Assessment of image quality and radiation dose in low-dose, low-contrast computed tomography coronary angiogram: a systematic review and meta-analysis.Clinical radiology · 2026