Real-time intraocular pressure monitoring and responsive drug release in preclinical models by an all-polymer smart contact lens
Yuting Cai, C B Zhang, Pengrui Dang, Chenshu Liu, Yuxuan Du, Reihaneh Haghniaz, Jiechen Wang, Safoora Khosravi,
Hong Kong University of Science and Technology · Sun Yat-sen University
Key points
From the paper's conclusion- Here, we report a battery-free, all-polymer microfluidic theranostic smart contact lens (AP-TSCL) that enables autonomous IOP-responsive glaucoma therapy without bulky electronic components.
Abstract
Glaucoma remains a leading cause of irreversible blindness worldwide, yet conventional topical therapies are often hampered by poor patient adherence, and standard in-office applanation tonometry captures only static measurements, failing to capture the dynamic nature of intraocular pressure (IOP). Smart contact lenses have emerged as promising tools for home-based, real-time IOP monitoring and feedback-guided therapy; however, most rely on bulky or rigid electronics that impair comfort, safety, and visual performance. Here, we report a battery-free, all-polymer microfluidic theranostic smart contact lens (AP-TSCL) that enables autonomous IOP-responsive glaucoma therapy without bulky electronic components. The AP-TSCL integrates a noninvasive microfluidic IOP sensor with a multistage, pressure-gated drug delivery architecture. Embedded microchannels define distinct activation thresholds, enabling lens deformation under elevated IOP to drive staged release from multiple drug reservoirs. Across in vitro (artificial eye model), ex vivo (enucleated bovine eyes), and in vivo (rabbit ocular hypertension) studies, the device achieved tonometry-aligned IOP tracking, pressure-triggered delivery of timolol or brimonidine above preset thresholds, and IOP lowering comparable to conventional topical therapy. This preclinical evidence suggests that the fully integrated platform may overcome key limitations of existing technologies and offers a clinically translatable solution for personalized ocular care.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Ultra-sensitive real-time monitoring of intraocular pressure with an integrated smart contact lens using parity-time symmetry wireless technologynpj Flexible Electronics · 2026
- Wearable Contact Lenses for Advanced Ocular Drug Delivery: Progress and ChallengesAdvanced Functional Materials · 2025
- Stretchable Wrinkled‐Gold Antenna for High‐Sensitivity Intraocular Pressure Monitoring Smart Contact LensAdvanced Materials Technologies · 2026
- Contact lens with stretchable distributed-feedback laser for intraocular pressure monitoringarXiv · 2026
- Flexible Transparent Capacitive Pressure Sensors for Multimodal Ophthalmic Monitoring: Intraocular Pressure Assessment, Movement Disorder Diagnosis, and Eye‐Tracking Human–Computer Interaction SystemAdvanced Functional Materials · 2025
- Beyond Sensitivity: Materials and Biointerface Design Rules for Aptamer-Enabled Smart Contact Lenses.ACS materials Au · 2026