Inter‐Device Consistency Between iFalcon and iCare in Mechanically Pressurized Ex Vivo Rat Eyes

Haobo Zhao, Yunhao Su, Lihui Yang, Ji Fu

Key points

From the paper's conclusion
  • This study has limitations: inherent errors from iCare TonoLab cannot be excluded, and mechanical compression precludes testing at extremely low IOP (0–5 mmHg).

Abstract

Objective To validate the accuracy of the iFalcon V700 rebound tonometer in measuring intraocular pressure (IOP) in enucleated normal rat eyes. Animals Studied Ten freshly enucleated normal rat eyes were tested within 12 h postmortem. Procedures Enucleated rat eyes were fixed to the front of the device. IOP measurements started at 5 mmHg and were gradually increased to 95 mmHg by mechanical pressure from the vitreous side toward the anterior chamber. The iFalcon V700 rebound tonometer measured central corneal IOP in rodent mode, and results were compared with the iCare TonoLab using linear regression analysis and Bland–Altman plots. Results Measurements from the iFalcon V700 correlated strongly with the iCare TonoLab ( r 2 = 0.98). Bland–Altman analysis showed a mean bias of −1.56 mmHg, with 95% limits of agreement from −6.71 to +3.59 mmHg, and slight underestimation in the moderate IOP range. In the 5–10 mmHg range, multiple attempts were needed to get valid readings with the iFalcon V700. Conclusions The iFalcon V700 enables accurate central‐corneal IOP measurement (5–95 mmHg) in enucleated normal rat eyes and shows good agreement with iCare TonoLab; repeated measurements at identical pressure levels improve precision. This study has limitations: inherent errors from iCare TonoLab cannot be excluded, and mechanical compression precludes testing at extremely low IOP (0–5 mmHg). Despite its promising potential for rodent tonometry, further systematic in vivo validation is required prior to its widespread, reliable use in rat ophthalmic research.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Inter‐Device Consistency Between iFalcon and iCare in Mechanically Pressurized Ex Vivo Rat Eyes — Science Skim