A Method for Measuring Intraocular Pressure in Wild-Type Zebrafish Using Rebound Tonometry: Characterization, Sex Differences, and Long-Term Repeatability.
Hsin-Lin Cheng, Pei-Hsuan Chou, Yung-Chang Yen, Jiann-Jou Yang
Department of BioMedical Sciences, Chung Shan Medical University, Taichung, Taiwan. · Department of Ophthalmology, Chi-Mei Hospital Liou-Ying, Tainan, Taiwan.
Key points
From the paper's conclusion- These findings indicated that our iCare tonometer-based experimental methods provide high measurement and time-point consistency, confirming their long-term reproducibility and reliability in measuring IOP in the zebrafish.
Abstract
Intraocular pressure (IOP) monitoring is a critical aspect of diagnosis, and management of ocular diseases. Therefore, a tractable vertebrate model capable of reliable IOP characterization is highly desirable. The present study aims to perform a preliminary investigation of a handheld rebound tonometer to stably measure IOP and further characterize baseline IOP values in the zebrafish model. The IOP in both eyes (oculus dexter, OD, and oculus sinister, OS) of wild-type zebrafish was measured using the iCare PRO rebound tonometer at the same time points on testing days for 18 months. The long-term IOP values of male zebrafish (3-21 months) were range from 21.96 ± 0.99 to 23.39 ± 0.32 mm Hg (OD) and 22.10 ± 0.93 to 23.38 ± 0.26 mm Hg (OS), while insignificant difference was noted among testing days (p > 0.05). Similarly, insignificant difference was observed in female zebrafish with IOP values range from 22.10 ± 1.12 to 23.19 ± 0.72 mm Hg (OD) and 22.30 ± 1.00 to 23.15 ± 0.36 mm Hg (OS) (p > 0.05). A comprehensive evaluation of IOP profiles confirmed no significant differences based on gender, time point, and between eyes. These findings indicated that our iCare tonometer-based experimental methods provide high measurement and time-point consistency, confirming their long-term reproducibility and reliability in measuring IOP in the zebrafish.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Inter‐Device Consistency Between iFalcon and iCare in Mechanically Pressurized Ex Vivo Rat EyesVeterinary Ophthalmology · 2026
- Changing tonometers, changing intraocular pressure? Rethinking intraocular pressure interpretation in dogs.The Journal of small animal practice · 2026
- Intraocular Pressure in Clinically Normal Alpacas: A Systematic Review and Meta-Analysis of Tono-Pen and TonoVet Tonometry.Animals : an open access journal from MDPI · 2026
- Multi-angle validation of the iFalcon V700 rebound tonometer against implantable pressure sensor in ex-vivo rabbit eyesScientific Reports · 2026
- Comparison of Horizontal and Vertical Tonovet Plus Clinical Measurement Configurations for Intraocular Pressure Assessment in Conscious CatsVeterinary Ophthalmology · 2026
- Standardized Ophthalmic Examination Protocol For Fish Developed During Assessment of 274 Eyes in Rockfish (Family Sebastidae)Veterinary Ophthalmology · 2026