Analytical Validation of β-Hydroxybutyrate Quantification in Feline Serum with the BT3500 Analyzer.
Alessandro Tirolo, Maria Chiara Sabetti, Alicia Maria Carrillo Heredero, Francesca Fidanzio, Andrea Corsini
Department of Veterinary Science, University of Parma, Parma, Italy.
Key points
From the paper's conclusion- The Randox β-OHB assay on the BT3500 demonstrated acceptable analytical performance in feline serum and is suitable for routine clinical use; frozen storage beyond 180 days may impact result accuracy.
Abstract
BACKGROUND: Quantitative measurement of serum β-hydroxybutyrate (β-OHB) is essential for detecting and monitoring ketosis and ketoacidosis in diabetic cats, particularly in the era of sodium-glucose cotransporter two (SGLT2) inhibitors, where euglycemic ketoacidosis may occur. The BT3500 analyzer has not been analytically validated for β-OHB quantification in feline serum. OBJECTIVES: Analytically validate the Randox Ranbut enzymatic β-OHB assay on feline serum using the BT3500 chemistry analyzer, including evaluation of sample stability under refrigerated and frozen storage conditions. METHODS: Residual feline serum samples and pooled sera were used to assess intra-assay and inter-assay precision, spike-recovery accuracy across five concentration increments, linearity under serial dilution, and analytical sensitivity (LoB, LoD, LoQ). Short-term stability was evaluated at 4°C for 7 days and long-term stability at -20°C at predefined time points up to 360 days. RESULTS: Intra-assay coefficient of variation (CV%) ranged from 1.68% to 14.48% and inter-assay CV% from 3.08% to 10.20%, with higher imprecision at low concentrations. Recoveries ranged from 110% to 122% (mean 114% ± 4%). Linearity was excellent (slope 1.012; R2 ≥ 0.9998). LoB, LoD, and LoQ were 0.044, 0.091, and 0.150 mmol/L, respectively. β-OHB was stable at 4°C up to 6 days and at -20°C up to 180 days, while decreases exceeding the acceptance limit were observed at 360 days. CONCLUSIONS: The Randox β-OHB assay on the BT3500 demonstrated acceptable analytical performance in feline serum and is suitable for routine clinical use; frozen storage beyond 180 days may impact result accuracy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Stability of beta-hydroxybutyrate in venous whole blood.Annals of clinical biochemistry · 2026
- A-386 Evaluation of near patient beta hydroxybutyrate testing in post mortem samples using the StatStrip analyserClinical Chemistry · 2026
- Whole-blood metabolic biomarkers improve detection of subclinical ketosis in postpartum grazing dairy cows.JDS communications · 2026
- Comparing clinical laboratory beta-hydroxybutyrate testing of vitreous humor against toxicological blood analyses in ketoacidosis deaths.Journal of forensic sciences · 2026
- Evaluation of a canine hemoglobin A1c point-of-care assay: analytical validation, clinical utility, and establishing a population-based reference interval.Journal of veterinary internal medicine · 2026
- Implementing point-of-care ketone testing and a 'three zone' ketone interpretation framework to predict dietary preparation and myocardial suppression in FDG PET for cardiac inflammationEuropean Heart Journal · 2026