Evaluation of a canine hemoglobin A1c point-of-care assay: analytical validation, clinical utility, and establishing a population-based reference interval.
Denzel Adam, Nicholas Noto, Michael F Rosser, Sarah Connolly, Arnon Gal
Department of Veterinary Clinical Medicine, College of Veterinary Medicine, University of Illinois at Urbana-Champaign, Urbana, IL 61802, United States.
Key points
From the paper's conclusion- Average intra- and inter-assay coefficients of variation were 3.92% and 8.47%. Linearity over the clinical range (0.86%-6.27% HbA1c%) was strong (R2 = 0.973) with minor constant bias. HbA1c remained stable for 5 days at 4 °C and up to 3 months at both -20 and -80 °C, but not reliably at 6 months. Lipemia and icterus exhibited variable interfering effects. Method comparison indicated non-interchangeability with the comparator assay. Diabetic dogs had significantly higher HbA1c than NDM (mean 5.10% vs 1.58%). ROC analysis yielded an AUC of 0.969, with a cutoff > 3.22% achieving 90.20% sensitivity and 100% specificity. The positive likelihood ratio was not estimable, and the odds ratio was 22.90 (95% CI, 3.60-146.90). Healthy young dogs' reference interval was 1.25% (90% CI, 1.19-1.59) to 2.75% (90% CI, 2.56-2.90).
Abstract
BACKGROUND: Hemoglobin A1c (HbA1c) assay validation is essential for establishing analytical reliability, evaluating discrimination between diabetic and non-diabetic (NDM) dogs, and establishing reference intervals. HYPOTHESIS/OBJECTIVES: Validate a canine HbA1c point-of-care analyzer (Vet Chroma™), evaluate its diagnostic performance, and establish a reference interval in healthy dogs. ANIMALS: Fifty-one diabetic, 44 NDM controls, and 42 healthy client-owned dogs. METHODS: Prospective validation of precision, linearity, interference, and storage stability. Method comparison used an immunoturbidimetric comparator assay (Olympus DxC 700 AU). Diagnostic performance was evaluated via receiver operating characteristic (ROC) analysis, likelihood ratios, and odds ratios. RESULTS: Average intra- and inter-assay coefficients of variation were 3.92% and 8.47%. Linearity over the clinical range (0.86%-6.27% HbA1c%) was strong (R2 = 0.973) with minor constant bias. HbA1c remained stable for 5 days at 4 °C and up to 3 months at both -20 and -80 °C, but not reliably at 6 months. Lipemia and icterus exhibited variable interfering effects. Method comparison indicated non-interchangeability with the comparator assay. Diabetic dogs had significantly higher HbA1c than NDM (mean 5.10% vs 1.58%). ROC analysis yielded an AUC of 0.969, with a cutoff > 3.22% achieving 90.20% sensitivity and 100% specificity. The positive likelihood ratio was not estimable, and the odds ratio was 22.90 (95% CI, 3.60-146.90). Healthy young dogs' reference interval was 1.25% (90% CI, 1.19-1.59) to 2.75% (90% CI, 2.56-2.90). CONCLUSIONS AND CLINICAL IMPORTANCE: The Vet Chroma™ provides reliable HbA1c measurements in canine whole blood and effectively discriminates diabetic from non-diabetic dogs. Results should be interpreted alongside clinical and clinicopathologic findings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of the Lifotronic H100 Plus analyzer for HbA1c determination.Scandinavian journal of clinical and laboratory investigation · 2026
- B-011 Evaluation of the Tosoh HLC-723GR01 HbA1c Analyzer Using Normal Hemoglobin Samples and in the Presence of Common Hemoglobin and Thalassemic VariantsClinical Chemistry · 2026
- B-005 Analytical Performance Evaluation of Two HbA1c Testing Methods on Roche cobas c513 AnalyzersClinical Chemistry · 2026
- B-059 Precision and Bias Verification of HbA1c by HPLC Following CLSI EP15-A3 in a Hospital Laboratory in Manabí, EcuadorClinical Chemistry · 2026
- An improved approach to modelling patient reclassification: HbA 1c as an exampleClinical Chemistry and Laboratory Medicine (CCLM) · 2026
- A-367 Assessment of Hemolysis Interference on HbA1c Measurement Using the Siemens Healthineers DCA Vantage AnalyzerClinical Chemistry · 2026