Leadless Pacemaker AVEIR VR: Performance of the Rate-Responsive Sensor During Exercise.
Valentine Gallet, Arnaud Denis, Olivier Césari, Sylvain Reuter, Rémi Chauvel
Clinique Saint Augustin, Bordeaux, France.
Key points
From the paper's conclusion- Exercise testing in patients implanted with an AVEIR VR leadless pacemaker demonstrated an overall appropriate rate-response, although with a non-negligible sensor latency. A substantial proportion of patients required post-exercise optimization of rate-response programming, highlighting the value of exercise-guided device programming to improve physiological chronotropic adaptation.
Abstract
BACKGROUND: To date, no published data have evaluated the exercise performance of the AVEIR VR leadless pacemaker. OBJECTIVE: We aimed to assess, during bicycle ergometer exercise testing, the relationship between the intrinsic sinus heart rate and the ventricular rate delivered by the AVEIR VR pacemaker. Patients with atrial fibrillation or those unable to perform exercise testing were excluded. METHODS: Patients underwent a bicycle ergometer exercise test using a workload protocol normalized to body weight (W/kg) according to their level of physical activity (sedentary or active). Simultaneous device interrogation was performed throughout the test. Intrinsic sinus heart rate and paced heart rate (or the projected paced heart rate in non-pacemaker-dependent patients) were recorded. Sensor latency was defined as the time required to achieve a 10-beats/min increase in paced heart rate above baseline. RESULTS: Between January and October 2025, 13 patients were included (69.9 ± 12.7 years; 23% women). Five patients were physically active, and eight patients were pacemaker-dependent. The mean sensor latency was 3.5 min in pacemaker-dependent patients and 2.7 min in non-pacemaker-dependent patients. Rate-response gain required adjustment in 38.5% of patients following exercise testing. Base rate and/or the maximum heart rate response were modified in 76.9% of patients. CONCLUSION: Exercise testing in patients implanted with an AVEIR VR leadless pacemaker demonstrated an overall appropriate rate-response, although with a non-negligible sensor latency. A substantial proportion of patients required post-exercise optimization of rate-response programming, highlighting the value of exercise-guided device programming to improve physiological chronotropic adaptation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Temperature Sensor-Mediated Intraoperative Tachycardia in a Leadless Pacemaker During Cardiac Surgery.JACC. Case reports · 2026
- Leadless Versus Transvenous Single‐Chamber Ventricular Pacemakers: Real‐World Evidence From Aveir VR Coverage With Evidence Development StudyJournal of the American Heart Association · 2025
- Multicenter Study of the Electrical Characteristics and Short-Term Outcomes of the Aveir VR Leadless Pacemaker.Journal of the American Heart Association · 2026
- Rate response of temperature-based sensors in leadless pacemakers during exercise test: a small cohort studyEuropean Heart Journal · 2026
- Multicenter study on electrical performance and predictive indicators in aveir VR pacemaker recipients.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026
- Leadless Pacemaker, Right Ventricular Pacing Burden, and Heart Failure.Heart rhythm · 2026