◆ Journal of the American Heart Association2026-09-18

Multicenter Study of the Electrical Characteristics and Short-Term Outcomes of the Aveir VR Leadless Pacemaker.

Jian-du Yang, Ruo-Gu Li, Xing-Bin Liu, Xiao-Lin Xue, Jiang-Hua Zhang, Yong-Mei Hu, Bao-Jian Zhang, Lin Tong, Hui Luo, Min Shen, Zhong-Li Chen, Xiahenazi Aiyasiding, Meng-Xing Cai, Xiaojian Chi, Yan Dai, Baopeng Tang, Ke-Ping Chen

一句话结论 · In one sentence

During Aveir implantation, commanded electrogram variations guide site localization. Initial COI increases (0-0.5 turns) are linked to optimal short-term thresholds. Monitoring impedance increase is vital, as a threshold of 230 Ω serves as a key indicator of device stability and fixation quality.

原始摘要(英文原文)· Original abstract
BACKGROUND: The Aveir leadless pacemaker uses an active fixation method, enabling real-time monitoring of electrical parameters during implantation. However, comprehensive studies on the electrical parameters during this procedure are rare. This study aims to analyze the electrical characteristics to further guide the implantation strategy and improve device stability and safety. METHODS: This multicenter retrospective study enrolled 119 patients (mean age, 70.18 years; 59.58% women) who received the Aveir VR leadless pacemaker from November 2024 to May 2025 at 10 centers in China. Intraprocedural variations in commanded electrogram, current of injury (COI), impedance, pacing threshold, and sensing parameters were meticulously documented. RESULTS: Commanded electrogram mapping demonstrated various morphologies (R, RS, QR, QRS, and QS), aiding localization. During fixation, 58.82% of patients exhibited an increased COI from mapping to 0.5 turns, which was associated with reduced short-term pacing thresholds. From 0.5 to 1 turn, 52.94% showed a further increase in COI. Receiver operating characteristic analysis revealed that an impedance increase has predictive value for short-term pacing thresholds, with an area under the receiver operating characteristic curve of 0.634 and a cutoff value of 230 Ω (sensitivity, 0.622; specificity, 0.41). Lead stability showed a moderate correlation with impedance increase (R=0.44, P<0.001), while the correlation with COI was weak. CONCLUSIONS: During Aveir implantation, commanded electrogram variations guide site localization. Initial COI increases (0-0.5 turns) are linked to optimal short-term thresholds. Monitoring impedance increase is vital, as a threshold of 230 Ω serves as a key indicator of device stability and fixation quality.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Multicenter Study of the Electrical Characteristics and Short-Term Outcomes of the Aveir VR Leadless Pacemaker. — 科研速览 Science Skim