Rate response of temperature-based sensors in leadless pacemakers during exercise test: a small cohort study
A Velraeds, K M Kooiman, F V Y Tjong, L R A Olde Nordkamp, R E Knops
Key points
From the paper's conclusion- In 2 patients (#2, #3), 33%, a delayed sensor response was observed, with the highest HR occurring during the recovery phase rather than during peak exercise.
Abstract
Background Leadless pacemakers (LPs) rely on rate-responsive sensors for pacing during physical activity in patients with chronotropic incompetence or pacing-dependency. Previous studies evaluated the rate-responsiveness of motion-based sensors, used in tined-fixation LPs. However, for the temperature-based sensor (TBS), used in screw-fixation LPs, the existing literature is limited. Evaluating the performance of TBSs is clinically relevant to ensure appropriate heart rate (HR) adaptation during exercise. TBSs detect temperature changes in the central venous blood pool. An increase in temperature is interpreted as physical activity; the LP adjusts the pacing rate accordingly. Optimization of the TBS can be done by adjusting the gain setting or the maximum sensor rate (MSR). Objective The aim of this study is to assess the performance of the TBS during exercise in patients with a LP. Methods Patients with a screw-fixation LP and an activated TBS underwent an exercise test (ET) on a bicycle, where an ECG together with the performed workload (WL) was registered during the ET. Results Seven patients with a LP and activated TBS underwent an ET: four patients were implanted with a single atrial LP, two with a single ventricular LP, and one with a dual chamber LP. Six patients had a paced heart rhythm during the ET. Figure 1 illustrates the rate-responsive performance of the TBS in their LP. Pacemaker settings prior to the ET were nominal and not individually optimized except for patient 6, in whom settings were based on previous transvenous pacemaker parameters and a walking test. In 2 patients (#2, #3), 33%, a delayed sensor response was observed, with the highest HR occurring during the recovery phase rather than during peak exercise. In patient 2, the sensor gain was increased after the ET, which resulted in a reduction of symptoms. For patient 3, no adjustments were made, as they remained asymptomatic. In all patients, the HR remained elevated during the initial post-ET period, reflecting a physiological response that is also observed in individuals without pacemaker therapy. Conclusion The TBS used in the screw-fixation LP for rate-responsive pacing, demonstrates an increase in heart rate during exercise testing in all patients. In patients with a delayed peak HR response, optimization was achieved through reprogramming.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Temperature Sensor-Mediated Intraoperative Tachycardia in a Leadless Pacemaker During Cardiac Surgery.JACC. Case reports · 2026
- Cardiopulmonary Exercise Testing in Patients with Cardiac Pacemakers: From Device Optimization to Individualized Cardiac Rehabilitation.Journal of clinical medicine · 2026
- Leadless Pacemaker AVEIR VR: Performance of the Rate-Responsive Sensor During Exercise.Pacing and clinical electrophysiology : PACE · 2026
- Sensor-Driven Pacemaker Tachycardia During Chest Physiotherapy: The Hidden Accelerometer Sensor Response in a Leadless Pacemaker in the VDD Mode.Journal of cardiovascular electrophysiology · 2026
- Pacing Performance of a Modular Communicative Leadless Pacing-Defibrillator System.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2026
- A nationwide DRG-based analysis of leadless pacemaker implantation in germany: temporal trends, patient characteristics, and in-hospital outcomesEuropean Heart Journal · 2026