Changes in Lead Parameters of Cardiac Implantable Electronic Devices After VT/VF Ablation Using a Dual-Energy Lattice-Tip Catheter.
Kazuto Hayasaka, Petr Peichl, Nicoletta Ventrella, Ibrahim Fahmawi, Josef Marek, Peter Štiavnický, Jana Hašková, Predrag Stojadinović,
Department of Cardiology, Institute for Clinical and Experimental Medicine, Prague, Czechia.
Key points
From the paper's conclusion- Marked acute pacing-threshold elevation was observed in both RV and LV leads after ventricular arrhythmia ablation using a dual-energy lattice-tip catheter. However, both RV and LV thresholds were no longer significantly different from baseline at 3 months. These findings suggest the importance of considering the anatomical relationship between ablation sites and CIED leads and of careful peri-procedural assessment of lead function during ventricular ablation.
Abstract
BACKGROUND: Interactions between cardiac implantable electronic devices (CIEDs) and pulsed field ablation (PFA) have been studied mainly in the context of atrial ablation, whereas the effects of ventricular ablation using a dual-energy lattice-tip catheter on ventricular leads remain unclear. METHODS: This retrospective study involved 79 patients (mean age 60.9 ± 13.6 years) with an implantable cardioverter-defibrillator or cardiac resynchronization therapy defibrillator who underwent ventricular ablation using a lattice-tip catheter between February 2024 and November 2025. Pacing thresholds were assessed before, immediately after, and 3 months after ablation and standardized to a pulse width of 0.4 ms using the Lapicque-Weiss model. Marked acute threshold elevation was defined as an increase in pacing threshold of ≥ 1.0 V immediately after ablation. The anatomical relationship between ablation sites and ventricular lead locations was assessed based on the American Heart Association 17-segment model of the left ventricle and classified as corresponding when ablation involved the same or opposing ventricular region as the lead location, adjacent when ablation involved a directly adjacent region, and remote otherwise. RESULTS: Marked acute pacing-threshold elevation occurred in 5 of 79 right ventricular (RV) leads (6.3%) and 2 of 22 evaluable left ventricular (LV) leads (9.1%). For RV leads, marked threshold elevation occurred in 4 of 14 patients (28.6%) in the corresponding group, 1 of 19 (5.3%) in the adjacent group, and none of 46 (0%) in the remote group. When threshold change was analyzed as a continuous variable, the acute increase was significantly greater in the corresponding group (+ 0.74 V; 95%CI 0.465-1.018) than in the adjacent (+ 0.31 V; 95%CI 0.076-0.550) and remote (+ 0.007 V; 95%CI - 0.145-0.160) groups. For LV leads, the pacing threshold increased significantly after ablation involving the corresponding segment (+ 0.71 V; 95%CI 0.167-1.245), although the overall association between anatomical relationship and threshold change was not significant. At 3 months, pacing thresholds in the corresponding groups were not significantly different from baseline for either RV or LV leads. CONCLUSION: Marked acute pacing-threshold elevation was observed in both RV and LV leads after ventricular arrhythmia ablation using a dual-energy lattice-tip catheter. However, both RV and LV thresholds were no longer significantly different from baseline at 3 months. These findings suggest the importance of considering the anatomical relationship between ablation sites and CIED leads and of careful peri-procedural assessment of lead function during ventricular ablation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effects of Dual-Energy Lattice-Tip Catheter Ablation of Ventricular Arrhythmias on Cardiac Implantable Electronic Device Performance.JACC. Clinical electrophysiology · 2026
- Permanent damage to implantable cardioverter-defibrillators during left-sided septal ventricular tachycardia ablation using a lattice-tip catheter: A case seriesHeart Rhythm · 2026
- Inadvertent ventricular fibrillation induced by radiofrequency ablation with the lattice-tip catheter in proximity to implantable cardioverter-defibrillator leadsHeart Rhythm · 2025
- Ventricular Fibrillation During Pulsed Field Ablation With Dual-Energy Focal Catheters in Patients With Cardiac Implantable Electronic Devices.Heart rhythm · 2026
- Context-dependent interactions between variable-loop pulsed field ablation and cardiac implantable electronic devices: a controlled preclinical study.European heart journal open · 2026
- Device malfunction following epicardial pulsed field ablation using a focal dual energy ablation catheter in a patient with cardiac resynchronization therapy defibrillator: a case report.European heart journal. Case reports · 2026