Effects of Dual-Energy Lattice-Tip Catheter Ablation of Ventricular Arrhythmias on Cardiac Implantable Electronic Device Performance.
Ehsan Mahmoodi, Arwa Younis, Koji Higuchi, Jakub Sroubek, Justin Z Lee, Roy Chung, Osama Dasa, Shady Nakhla,
Department of Cardiovascular Medicine, Cardiac Electrophysiology Section, Cleveland Clinic, Cleveland, Ohio, USA. · Department of Cardiovascular Medicine, Cardiac Electrophysiology Section, Cleveland Clinic, Cleveland, Ohio, USA. Electronic address: santanp3@ccf.org.
Key points
From the paper's conclusion- Ventricular arrhythmia ablation using the Sphere-9 dual-energy lattice-tip catheter was associated with small but persistent changes in right and left ventricular lead parameters. Serious CIED interactions were uncommon, and further studies are needed to better understand the underlying mechanisms.
Abstract
BACKGROUND: The Sphere-9 lattice-tip catheter (Affera, Medtronic) enables delivery of both radiofrequency ablation (RFA) and pulsed field ablation (PFA) for ventricular arrhythmia (VA) ablation. Data on its impact on cardiac implantable electronic devices (CIEDs) are limited. OBJECTIVES: This study sought to evaluate changes in CIED function following VA ablation using the Sphere-9 catheter. METHODS: We reviewed 65 consecutive patients (median age 71 years, 94% male, 63% ischemic cardiomyopathy, median left ventricular ejection fraction 31%) with pre-existing CIEDs who underwent VA ablation (62 ventricular tachycardia [95%], 3 premature ventricular contraction [5%]) using the Sphere-9 catheter (combined RFA/PFA (53 [82%]), RFA alone (6 [9%]), PFA alone (6 [9%])). Device interrogation was performed within 30 days preprocedure, within 7 days postprocedure, and at 3-month (+/- 30 days) follow-up. RESULTS: At 3-months, right ventricular leads demonstrated reduced sensing amplitude (11.7-10.0 mV, P = 0.04), pacing impedance (441-422 Ω, P < 0.001), shock impedance (64-59 Ω, P = 0.01), and increased capture threshold (0.9-1.0 V at 0.4-0.5 milliseconds, P = 0.03). Left ventricular (LV) leads demonstrated reduced pacing impedance (544-475 Ω, P = 0.04), whereas LV capture threshold remained elevated (1.0-1.4 V, P = 0.08). Right atrial lead parameters remained stable. No patient required CIED revision. One patient experienced cardiac resynchronization therapy-D pulse generator failure during PFA delivery 7.5 mm from an LV lead tip electrode, requiring generator replacement. CONCLUSIONS: Ventricular arrhythmia ablation using the Sphere-9 dual-energy lattice-tip catheter was associated with small but persistent changes in right and left ventricular lead parameters. Serious CIED interactions were uncommon, and further studies are needed to better understand the underlying mechanisms.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Permanent damage to implantable cardioverter-defibrillators during left-sided septal ventricular tachycardia ablation using a lattice-tip catheter: A case seriesHeart Rhythm · 2026
- Changes in Lead Parameters of Cardiac Implantable Electronic Devices After VT/VF Ablation Using a Dual-Energy Lattice-Tip Catheter.Journal of cardiovascular electrophysiology · 2026
- Inadvertent ventricular fibrillation induced by radiofrequency ablation with the lattice-tip catheter in proximity to implantable cardioverter-defibrillator leadsHeart Rhythm · 2025
- Real‐World Adverse Events Associated With the Sphere‐9 Dual‐Energy Arrhythmia Ablation CatheterJournal of Cardiovascular Electrophysiology · 2026
- Ventricular Fibrillation During Pulsed Field Ablation With Dual-Energy Focal Catheters in Patients With Cardiac Implantable Electronic Devices.Heart rhythm · 2026
- Pulsed field ablation of atrial arrhythmias in patients with cardiac implantable electronic devices: a multicenter evaluation of electrical integrity and device safety.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026