Device malfunction following epicardial pulsed field ablation using a focal dual energy ablation catheter in a patient with cardiac resynchronization therapy defibrillator: a case report.
Rahin Wahedi, Julia Vogler, Stephan Willems, Arian Sultan
Department of Cardiology and Internal Intensive Care, Asklepios Klinik St. Georg, Lohmühlenstraße 5, 20099 Hamburg, Germany.
Key points
From the paper's conclusion- To our knowledge, this is the first report of a clinically relevant transient CIED malfunction temporally associated with epicardial ventricular PFA using a focal single-tip dual-energy ablation catheter.
Abstract
BACKGROUND: Pulsed field ablation (PFA) is increasingly used, but experience with ventricular, and particularly epicardial, applications in patients with cardiac implantable electronic devices (CIEDs) remains limited. CASE SUMMARY: A 49-year-old woman with LMNA- and SCN5A-associated cardiomyopathy, mildly reduced left ventricular ejection fraction (45%), and a cardiac resynchronization therapy defibrillator (CRT-D) presented with recurrent ventricular tachycardia (VT) and implantable cardioverter-defibrillator shocks. Previous endocardial VT ablation had revealed no relevant endocardial low-voltage substrate.Repeat ablation was performed using a combined endo-epicardial approach. Epicardial mapping identified extensive low-voltage areas, with the best pace-map match at the basal lateral left ventricle. Ablation was performed using a focal dual-energy catheter with radiofrequency energy up to 40 W and PFA (2400 V).Following epicardial PFA in proximity to the device generator and left ventricular lead, pacing abruptly changed from VVI 30/min to DDD 70/min. Initial attempts at device interrogation failed. After communication was restored, interrogation demonstrated that the CRT-D had reverted to factory settings. The device was successfully reset and reprogrammed. Further PFA near the LV lead or generator was avoided. Sensing, pacing thresholds, and lead impedances were preserved. DISCUSSION: To our knowledge, this is the first report of a clinically relevant transient CIED malfunction temporally associated with epicardial ventricular PFA using a focal single-tip dual-energy ablation catheter. Although no permanent generator or lead dysfunction occurred, this observation highlights the potential for PFA-CIED interactions beyond conventional PFA workflows and across different catheter technologies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Ventricular Fibrillation During Pulsed Field Ablation With Dual-Energy Focal Catheters in Patients With Cardiac Implantable Electronic Devices.Heart rhythm · 2026
- Pulsed field ablation of atrial arrhythmias in patients with cardiac implantable electronic devices: a multicenter evaluation of electrical integrity and device safety.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026
- Context-dependent interactions between variable-loop pulsed field ablation and cardiac implantable electronic devices: a controlled preclinical study.European heart journal open · 2026
- Pulsed Field Ablation in Patients With Cardiac Implantable Electric Devices: Vigilance Is WarrantedPacing and Clinical Electrophysiology · 2025
- Successful pulsed field ablation for refractory premature ventricular contractions using a novel dual-energy focal catheter: a case report.European heart journal. Case reports · 2026
- Endo-Epicardial Ventricular Tachycardia Ablation With Pulsed-Field Energy in an Extravascular ICD RecipientJACC Case Reports · 2026