The Correction Factor for Monitoring Efanesoctocog Alfa Is Not Always 2.5.
Sean Platton, Jude Platton, Katarzyna Kniola, Laura Aiken, Janine Collins, Suthesh Sivapalaratnam, Priyanka Raheja
The Royal London Hospital Haemophilia Centre, Barts Health NHS Trust, London, UK. · Haemostasis Laboratory, Royal London Hospital, Barts Health NHS Trust, London, UK.
Key points
From the paper's conclusion- A universal CF for Altuvoct assays may lead to clinically significant FVIII:C underestimation across the measurement range, and is inappropriate in most cases. Should product-specific calibrators (and controls) become available, laboratories should verify their assays for Altuvoct using them so that CF are not required.
Abstract
INTRODUCTION: Efanesoctocog alfa (Altuvoct) is a recombinant super-extended half-life factor (F)VIII concentrate licensed in the United Kingdom for patients with severe and moderate haemophilia A. Concentrate potency is assigned by one-stage assays (OSA) using Actin FSL. Chromogenic FVIII:C assays (CA) and some OSA are inaccurate for Altuvoct; the product information sheet suggests results of these be divided by 2.5 to approximate FVIII:C. UK experts recommend against using correction factors (CF), with guidance that only validated OSA should be used for Altuvoct, or, in the presence of FVIII-mimetics, that CA results are reported with a caveat about assay inaccuracy. METHODS: Over four months, we assessed the impact of a fixed CF on FVIII:C, testing 70 samples collected during pharmacokinetic assessment when patients switched to Altuvoct. All were tested using OSA (Actin FSL and Actin FS) and CA (BIOPHEN and Siemens FVIII:C) on a Sysmex CN-6500 analyser, calibrated against a plasma standard. RESULTS: For samples with Altuvoct between 10 and 269 IU/dL, CF increased as FVIII:C increased when using CA and decreased when using OSA (Actin FS), suggesting that a single CF across that range is inappropriate. CONCLUSIONS: A universal CF for Altuvoct assays may lead to clinically significant FVIII:C underestimation across the measurement range, and is inappropriate in most cases. Should product-specific calibrators (and controls) become available, laboratories should verify their assays for Altuvoct using them so that CF are not required.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The Laboratory Monitoring of Altuvoct (Efanesoctocog Alfa): Recommendations From the Laboratory Working Party of the United Kingdom Haemophilia Centres Doctor's OrganisationHaemophilia · 2025
- Assay of Efanesoctocog Alfa in 200 Centres: Data From Collaborative NEQAS BC (United Kingdom) and ECAT (Netherlands) Exercise Autumn 2024Haemophilia · 2025
- Assay or astray? discrepancies in efanesoctocog alfa monitoring.Research and practice in thrombosis and haemostasis · 2026
- Comparison of Actin FSL and Pathromtin SL aPTT Reagents in Patients Receiving Efanesoctocog Alfa: A Retrospective Laboratory Study.Diagnostics (Basel, Switzerland) · 2026
- Discrepancies in plasma alkaline phosphatase reruns on Cobas c503 Roche analysers: a retrospective analysis and impact of dilution.Diagnosis (Berlin, Germany) · 2026
- On the optimal time points for assessing the cortisol response in the high-dose ACTH stimulation test - a literature review.Journal of molecular endocrinology · 2026