Comparison of Actin FSL and Pathromtin SL aPTT Reagents in Patients Receiving Efanesoctocog Alfa: A Retrospective Laboratory Study.
Lena Weindl, Martin Olivieri, Bernd H Northoff, Michael Spannagl, Juergen Patzke, Michael Weigand, Daniel Teupser, Mathias Bruegel
Institute of Laboratory Medicine, LMU University Hospital, LMU Munich, 81377 Munich, Germany. · Pediatric Thrombosis and Haemostasis Unit, Dr. von Hauner Children's Hospital, LMU University Hospital, LMU Munich, 80337 Munich, Germany.
Key points
From the paper's conclusion- This study compared Actin FSL and Pathromtin SL in post-infusion EFA samples, assessed lot-to-lot variability, and evaluated aPTT performance in detecting reduced FVIII activity.
Abstract
Background: Efanesoctocog alfa (EFA) is an extended half-life factor VIII (FVIII) concentrate. Although FVIII assay variability in EFA monitoring is established, data on the comparability of activated partial thromboplastin time (aPTT) reagents remain limited. This study compared Actin FSL and Pathromtin SL in post-infusion EFA samples, assessed lot-to-lot variability, and evaluated aPTT performance in detecting reduced FVIII activity. Methods: This retrospective, single-center study analyzed 106 paired samples from 53 patients receiving once-weekly EFA at trough level, peak level (1 h post-infusion), or follow-up (median 4.0 days). FVIII activity was measured using an Actin FSL-based one-stage clotting assay with FVIII-deficient plasma. aPTT was determined using Actin FSL and Pathromtin SL on Sysmex coagulation analyzers. Method comparison included Passing-Bablok regression and modified Bland-Altman analysis. Results: Pathromtin SL produced longer clotting times (mean bias: 5.2 s, mean relative difference: 13.1%). Both aPTT reagents showed a strong correlation in post-infusion EFA samples (Pearson's r = 0.878). No significant differences in regression slopes were observed across reagent lot combinations; maximum relative inter-lot difference was 6.5%. Using assay-specific cut-offs, both aPTT assays showed 100% sensitivity for detecting FVIII activity < 30% in this cohort. Nevertheless, prolonged aPTT values also occurred at FVIII activity ≥ 30%, indicating limited specificity. Conclusions: Both Actin FSL and Pathromtin SL aPTT assays identified all samples with Actin FSL-based FVIII activity < 30%. However, significantly longer clotting times with Pathromtin SL necessitate assay-specific interpretation in clinical practice. Findings should be interpreted considering the limited sample size, retrospective single-center design and potential methodological dependency.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Assay of Efanesoctocog Alfa in 200 Centres: Data From Collaborative NEQAS BC (United Kingdom) and ECAT (Netherlands) Exercise Autumn 2024Haemophilia · 2025
- Assay or astray? discrepancies in efanesoctocog alfa monitoring.Research and practice in thrombosis and haemostasis · 2026
- The Correction Factor for Monitoring Efanesoctocog Alfa Is Not Always 2.5.Haemophilia : the official journal of the World Federation of Hemophilia · 2026
- The Laboratory Monitoring of Altuvoct (Efanesoctocog Alfa): Recommendations From the Laboratory Working Party of the United Kingdom Haemophilia Centres Doctor's OrganisationHaemophilia · 2025
- Real-World Clinical Data and Thrombin Generation of Efanesoctocog Alfa Prophylaxis in Pediatric Patients With Haemophilia A.Haemophilia : the official journal of the World Federation of Hemophilia · 2026
- An In Vitro Study Simulating Prophylactic Anticoagulation in Haemophilia A and BeJHaem · 2026