Assay or astray? discrepancies in efanesoctocog alfa monitoring.
Agathe Herb, Jordan Wimmer, Sarra Derbal, Manon Dolt, Léa Pierre, Olivier Feugeas, Laurent Mauvieux, Dominique Desprez,
Laboratoire d'Hématologie, Unité Hémostase, Hôpitaux Universitaires de Strasbourg, Strasbourg, France. · Centre de Ressource et Compétences des Maladies Hémorragiques Congénitales, Hôpitaux Universitaires de Strasbourg, Strasbourg, France.
Key points
From the paper's conclusion- Assay performance for EFA differs between spiked and patient samples, underscoring the limitations of in vitro evaluations. No assay provided consistent agreement with the reference assay across the full FVIII range, highlighting the need for cautious interpretation and improved standardization of EFA monitoring in clinical practice.
Abstract
BACKGROUND: Efanesoctocog alfa (EFA) is an ultraextended half-life factor VIII (FVIII) with a unique structure that may influence laboratory monitoring. Although assay discrepancies have been described in spiked samples, data in patient samples are limited. OBJECTIVES: To assess the performance of seven FVIII assays for measuring EFA in patient plasma and compare results with those from spiked samples. METHODS: FVIII activity was measured using four one-stage assays (OSAs) and three chromogenic assays (CSAs) on two analyzers. Assays were calibrated either with human plasma or an EFA-specific calibrator. Measurements were performed in FVIII-deficient plasma spiked with EFA and in samples from patients with hemophilia A treated with EFA. Results were compared with a reference assay (Actin FSL) and expressed as recovery percentages (acceptable range: 80%-120%). RESULTS: In spiked samples, plasma-calibrated CSA overestimated FVIII activity (+37% to +94%), whereas EFA-calibrated CSA showed good agreement with the reference assay (+2% to +16%). OSA results were generally acceptable, with slight underestimation depending on the reagent. In patient samples, patterns differed markedly: most assays deviated significantly from the reference, and none remained within the acceptable recovery range across the full FVIII spectrum. Compared with spiked samples, discrepancies were more pronounced at later time points and trough levels, where OSA tended to overestimate FVIII activity, whereas EFA-calibrated CSA underestimated activity. CONCLUSION: Assay performance for EFA differs between spiked and patient samples, underscoring the limitations of in vitro evaluations. No assay provided consistent agreement with the reference assay across the full FVIII range, highlighting the need for cautious interpretation and improved standardization of EFA monitoring in clinical practice.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Assay of Efanesoctocog Alfa in 200 Centres: Data From Collaborative NEQAS BC (United Kingdom) and ECAT (Netherlands) Exercise Autumn 2024Haemophilia · 2025
- Comparison of Actin FSL and Pathromtin SL aPTT Reagents in Patients Receiving Efanesoctocog Alfa: A Retrospective Laboratory Study.Diagnostics (Basel, Switzerland) · 2026
- The Correction Factor for Monitoring Efanesoctocog Alfa Is Not Always 2.5.Haemophilia : the official journal of the World Federation of Hemophilia · 2026
- The Laboratory Monitoring of Altuvoct (Efanesoctocog Alfa): Recommendations From the Laboratory Working Party of the United Kingdom Haemophilia Centres Doctor's OrganisationHaemophilia · 2025
- Challenges in monitoring factor VIII activity during hemophilia a treatment: toward individualized monitoring strategies.Expert review of hematology · 2026
- A Modified Intrinsic Thrombin Generation Assay for Functional Assessment of Factor and Non-Factor Therapies in HemophiliaBiomedicines · 2026