On the optimal time points for assessing the cortisol response in the high-dose ACTH stimulation test - a literature review.
Karlijn Koops, Annenienke C van de Ven, Michiel N Kerstens, Judith Ma Emmen, Nienke R Biermasz, Annemieke C Heijboer, Jacquelien J Hillebrand
Department of Laboratory Medicine, Endocrine Laboratory, Amsterdam UMC, Vrije Universiteit Amsterdam, Boelelaan 1117, 1081 HV Amsterdam, The Netherlands. · Department of Internal Medicine, Division of Endocrinology, Radboud University Medical Center , Nijmegen, The Netherlands.
Key points
From the paper's conclusion- In conclusion, serum cortisol measurements at t30 and t60 during the high-dose ACTH stimulation test seem equally accurate.
Abstract
The high-dose ACTH stimulation test is used to assess adrenal insufficiency. Although the test has a rather simple design, many variables in pretest conditions, performing the test and cortisol methods impact the interpretation of the test result, for which rigid cut-offs were previously used. There is growing attention to standardized work-up and using well-standardized analytical methods and assay-specific cortisol cut-offs. There is less consensus on whether post-stimulation serum cortisol concentrations should be measured at 30 minutes (t30) or 60 minutes (t60) or both. Often the same cortisol cut-off is used for measurements at both time points, even though the continuous cortisol response in this time frame suggests using different cut-offs. We reviewed studies describing cortisol measurements at t30 and t60 during the high-dose AST and studied diagnostic accuracy. We noted that cortisol cut-offs should be specific to the assay used and, when using modern assays, lower than the original 500 or 550 nmol/L. When using a single cut-off for t30 and t60, measurements at the t60 time point showed higher specificity. However, when using time point-specific cortisol cut-offs, the specificity of cortisol measurements at t30 or t60 appeared comparable. In conclusion, serum cortisol measurements at t30 and t60 during the high-dose ACTH stimulation test seem equally accurate. A similar performance may be achieved using either time point alone, provided time point and assay-specific cut-offs are used. It should be noted, however, that the results of the high-dose ACTH stimulation test should always be interpreted within the clinical context.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Cortisol Cutoffs for Adrenal Insufficiency Using Elecsys II: A Systematic Review.Clinical endocrinology · 2026
- Salivary cortisol measurements for reliable numerical maximum concentration after the low dose 1 µg ACTH intravenously in healthy individualsScientific Reports · 2026
- Diagnostic accuracy of the ovine CRH test in ACTH-dependent Cushing syndrome.Endocrine connections · 2026
- Use of EQA for assessing variation in post-analytical interpretation of results using a clinical case of ?adrenal insufficiency taking into consideration variation in analytical performance of cortisol.British journal of biomedical science · 2026
- Evaluation of basal cortisol and the adrenocorticotropic hormone stimulation test in 982 dogs undergoing investigation for hypoadrenocorticism.Journal of veterinary internal medicine · 2026
- Evaluation of Desmopressin Stimulation Test Specificity in Healthy VolunteersEndocrine Connections · 2026