Device-related differences and interchangeability of forced expiratory volume in 1 s measurements: A comparison of three spirometers.
Leon L Csonka, Antti Tikkakoski, Jussi Karjalainen, Lauri Lehtimäki
Faculty of Medicine and Health Technology, Tampere University, Tampere, Finland. · Department of Clinical Physiology and Nuclear Medicine, Tampere University Hospital, Tampere, Finland.
Key points
From the paper's conclusion- Systematic bias and varying agreement between the laboratory spirometer and the handheld microspirometers indicate that FEV1 measurements obtained with different devices should not be assumed to be fully interchangeable. This does not imply that handheld microspirometry is inferior, but rather that device-specific characteristics must be considered when interpreting and comparing FEV1 values.
Abstract
BACKGROUND: Forced expiratory volume in 1 s (FEV1) obtained using different types of laboratory spirometers and handheld microspirometers is often used interchangeably in research, despite substantial differences between devices. The aim of this study was to assess device-related differences and the interchangeability of FEV1 measurements among three commonly used spirometers. METHODS: We recruited 20 healthy adults experienced in performing spirometry. Each participant performed three measurements with each of the three spirometers: the handheld microspirometers Medikro Duo and MIR Spirobank, and the laboratory spirometer Vyntus Pneumo. RESULTS: In the linear mixed-effects model, there was a significant main effect of device on FEV1 (F(2,38) = 25.25, p < 0.001), indicating that FEV1 values differed between devices. Spirobank yielded the highest FEV1 values, whereas the Medikro Duo yielded the lowest. The mean difference in FEV1 between Medikro Duo and Spirobank was -0.25 L (-0.34 to -0.16, p < 0.001), and between Medikro Duo and Vyntus Pneumo -0.16 L (-0.25 to -0.07, p < 0.001). The comparison between Spirobank and Vyntus Pneumo showed a smaller difference of 0.09 L (0.00 to 0.18), which narrowly missed statistical significance (p = 0.050). Bland-Altman plots showed wide limits of agreement between spirometers, suggesting poor interchangeability. CONCLUSIONS: Systematic bias and varying agreement between the laboratory spirometer and the handheld microspirometers indicate that FEV1 measurements obtained with different devices should not be assumed to be fully interchangeable. This does not imply that handheld microspirometry is inferior, but rather that device-specific characteristics must be considered when interpreting and comparing FEV1 values.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy and reliability of personal home spirometers: a multicentre laboratory quality-control study in children and adults.Respiratory medicine · 2026
- Comparison of two respiratory oscillometers in patients with a broad range of lung function patterns.Frontiers in allergy · 2026
- Agreement and Repeatability of Three Fractional Exhaled Nitric Oxide Analyzers in Patients with Chronic Obstructive Pulmonary Disease: A Prospective Comparative Study.Journal of clinical medicine · 2026
- Bench Agreement of Two Units of a Low-Cost Digital Manometers with a Clinical Respiratory Pressure Comparator: A Laboratory-Based Comparison.Diagnostics (Basel, Switzerland) · 2026
- The Impact of Inspiratory Muscle Function on Inhaler Device Adaptability in Stable COPD Patients.Journal of aerosol medicine and pulmonary drug delivery · 2026
- Diagnostic Accuracy of Tidal Breathing Patterns in Assessing Airflow Obstruction Compared with Spirometry.International journal of chronic obstructive pulmonary disease · 2026