Device-related differences and interchangeability of forced expiratory volume in 1 s measurements: A comparison of three spirometers.

Leon L Csonka, Antti Tikkakoski, Jussi Karjalainen, Lauri Lehtimäki

Faculty of Medicine and Health Technology, Tampere University, Tampere, Finland. · Department of Clinical Physiology and Nuclear Medicine, Tampere University Hospital, Tampere, Finland.

Key points

From the paper's conclusion
  • Systematic bias and varying agreement between the laboratory spirometer and the handheld microspirometers indicate that FEV1 measurements obtained with different devices should not be assumed to be fully interchangeable. This does not imply that handheld microspirometry is inferior, but rather that device-specific characteristics must be considered when interpreting and comparing FEV1 values.

Abstract

BACKGROUND: Forced expiratory volume in 1 s (FEV1) obtained using different types of laboratory spirometers and handheld microspirometers is often used interchangeably in research, despite substantial differences between devices. The aim of this study was to assess device-related differences and the interchangeability of FEV1 measurements among three commonly used spirometers. METHODS: We recruited 20 healthy adults experienced in performing spirometry. Each participant performed three measurements with each of the three spirometers: the handheld microspirometers Medikro Duo and MIR Spirobank, and the laboratory spirometer Vyntus Pneumo. RESULTS: In the linear mixed-effects model, there was a significant main effect of device on FEV1 (F(2,38) = 25.25, p < 0.001), indicating that FEV1 values differed between devices. Spirobank yielded the highest FEV1 values, whereas the Medikro Duo yielded the lowest. The mean difference in FEV1 between Medikro Duo and Spirobank was -0.25 L (-0.34 to -0.16, p < 0.001), and between Medikro Duo and Vyntus Pneumo -0.16 L (-0.25 to -0.07, p < 0.001). The comparison between Spirobank and Vyntus Pneumo showed a smaller difference of 0.09 L (0.00 to 0.18), which narrowly missed statistical significance (p = 0.050). Bland-Altman plots showed wide limits of agreement between spirometers, suggesting poor interchangeability. CONCLUSIONS: Systematic bias and varying agreement between the laboratory spirometer and the handheld microspirometers indicate that FEV1 measurements obtained with different devices should not be assumed to be fully interchangeable. This does not imply that handheld microspirometry is inferior, but rather that device-specific characteristics must be considered when interpreting and comparing FEV1 values.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Device-related differences and interchangeability of forced expiratory volume in 1 s measurements: A comparison of three spirometers — Science Skim