Diagnostics (Basel, Switzerland)

Bench Agreement of Two Units of a Low-Cost Digital Manometers with a Clinical Respiratory Pressure Comparator: A Laboratory-Based Comparison.

Rodrigo Torres-Castro, Matías Otto-Yáñez, Claudia Saldaña-Villegas, Luis Vasconcello-Castillo, Giovanna Binetti, Homero Puppo, Marisol Barros-Poblete, Roberto Vera-Uribe

Department of Physical Therapy, Faculty of Medicine, University of Chile, Santiago 8380453, Chile. · Grupo de Investigación en Salud, Funcionalidad y Actividad Física (GISFAF), Kinesiología, Facultad de Ciencias de la Salud, Universidad Autónoma de Chile, Santiago 7500912, Chile.

Key points

From the paper's conclusion
  • For MEP, both units showed higher readings than the comparator (mean differences 0.52–0.72 cmH2O), and differences increased with pressure; at 70 cmH2O under low compliance, the predicted upper limit of agreement reached 2.44–2.64 cmH2O.

Abstract

Background: Maximal inspiratory (MIP) and expiratory (MEP) pressures are widely used indices of respiratory muscle strength, but clinical manometers are costly. We evaluated the bench agreement of two units of a low-cost digital manometer with a clinical respiratory pressure comparator. Methods: Two BioPimo 101® units and a Vitalograph Pneumotrac 6800® were connected simultaneously to a mechanical insufflation-exsufflation device coupled to a test lung. Pressures of 5-70 cmH2O were applied in 5-cmH2O steps under two compliance conditions, in triplicate (56 conditions; 336 paired differences analyzed as repeated measurements). Agreement was assessed with Bland-Altman analysis, condition-clustered bootstrap confidence intervals, regression-based limits of agreement and ICC (2,1). Results: For MIP, mean differences were -0.03 to 0.21 cmH2O (limits of agreement -0.42 to 0.72 cmH2O). For MEP, both units showed higher readings than the comparator (mean differences 0.52-0.72 cmH2O), and differences increased with pressure; at 70 cmH2O under low compliance, the predicted upper limit of agreement reached 2.44-2.64 cmH2O. ICCs exceeded 0.999. Conclusions: Both units agreed closely with the comparator under bench conditions, particularly for MIP, whereas MEP showed positive, pressure-dependent differences. Without a traceable pressure standard and with only two units tested, these findings do not establish absolute accuracy or clinical interchangeability.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Bench Agreement of Two Units of a Low-Cost Digital Manometers with a Clinical Respiratory Pressure Comparator: A Laboratory-Based Comparison — Science Skim