Bench Agreement of Two Units of a Low-Cost Digital Manometers with a Clinical Respiratory Pressure Comparator: A Laboratory-Based Comparison.
Rodrigo Torres-Castro, Matías Otto-Yáñez, Claudia Saldaña-Villegas, Luis Vasconcello-Castillo, Giovanna Binetti, Homero Puppo, Marisol Barros-Poblete, Roberto Vera-Uribe
Department of Physical Therapy, Faculty of Medicine, University of Chile, Santiago 8380453, Chile. · Grupo de Investigación en Salud, Funcionalidad y Actividad Física (GISFAF), Kinesiología, Facultad de Ciencias de la Salud, Universidad Autónoma de Chile, Santiago 7500912, Chile.
Key points
From the paper's conclusion- For MEP, both units showed higher readings than the comparator (mean differences 0.52–0.72 cmH2O), and differences increased with pressure; at 70 cmH2O under low compliance, the predicted upper limit of agreement reached 2.44–2.64 cmH2O.
Abstract
Background: Maximal inspiratory (MIP) and expiratory (MEP) pressures are widely used indices of respiratory muscle strength, but clinical manometers are costly. We evaluated the bench agreement of two units of a low-cost digital manometer with a clinical respiratory pressure comparator. Methods: Two BioPimo 101® units and a Vitalograph Pneumotrac 6800® were connected simultaneously to a mechanical insufflation-exsufflation device coupled to a test lung. Pressures of 5-70 cmH2O were applied in 5-cmH2O steps under two compliance conditions, in triplicate (56 conditions; 336 paired differences analyzed as repeated measurements). Agreement was assessed with Bland-Altman analysis, condition-clustered bootstrap confidence intervals, regression-based limits of agreement and ICC (2,1). Results: For MIP, mean differences were -0.03 to 0.21 cmH2O (limits of agreement -0.42 to 0.72 cmH2O). For MEP, both units showed higher readings than the comparator (mean differences 0.52-0.72 cmH2O), and differences increased with pressure; at 70 cmH2O under low compliance, the predicted upper limit of agreement reached 2.44-2.64 cmH2O. ICCs exceeded 0.999. Conclusions: Both units agreed closely with the comparator under bench conditions, particularly for MIP, whereas MEP showed positive, pressure-dependent differences. Without a traceable pressure standard and with only two units tested, these findings do not establish absolute accuracy or clinical interchangeability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Individualized respiratory exercise device originated in Türkiye: design, prototype, and first resultsTURKISH JOURNAL OF MEDICAL SCIENCES · 2026
- A New Protocol for Evaluating Maximum Respiratory Pressures in Individuals With Multiple Sclerosis: Concurrent Validity and Test-Retest Reliability.Journal of neurologic physical therapy : JNPT · 2026
- The Impact of Inspiratory Muscle Function on Inhaler Device Adaptability in Stable COPD Patients.Journal of aerosol medicine and pulmonary drug delivery · 2026
- Automated Versus Manual Occlusion Pressure at 100 ms With Hamilton C6 Using the Software Version 1.2.3: From Bedside to the Bench and Back to Bedside.Critical care explorations · 2026
- Effect of the Measurement Interface on Maximal Expiratory Pressure During Voluntary Cough in Young Adults.Diagnostics (Basel, Switzerland) · 2026
- Accuracy and reliability of personal home spirometers: a multicentre laboratory quality-control study in children and adults.Respiratory medicine · 2026