Agreement and Repeatability of Three Fractional Exhaled Nitric Oxide Analyzers in Patients with Chronic Obstructive Pulmonary Disease: A Prospective Comparative Study.
Claudio Candia, Silvestro Ennio D'Anna, Antonio Iovine, Carmen Lombardi, Claudia Merola, Mauro Maniscalco
Istituti Clinici Scientifici Maugeri IRCCS, Pulmonary Rehabilitation Unit of Telese Terme Institute, 82037 Telese Terme, Italy.
Key points
From the paper's conclusion- Background and aim: Fractional exhaled nitric oxide (FeNO) is a non-invasive biomarker of airway inflammation that may provide complementary information on inflammatory phenotypes in chronic obstructive pulmonary disease (COPD).
Abstract
Background and aim: Fractional exhaled nitric oxide (FeNO) is a non-invasive biomarker of airway inflammation that may provide complementary information on inflammatory phenotypes in chronic obstructive pulmonary disease (COPD). The increasing availability of portable FeNO analyzers requires assessment of their reproducibility and agreement before measurements obtained with different devices can be considered interchangeable. In this study, we aimed to compare the performance of three FeNO analyzers in patients with COPD. Methods: In this prospective study, consecutive patients with COPD underwent FeNO assessment using three analyzers: Medisoft Hyp'Air FeNO, Sunvou CA2122, and Sunvou TM2120. Two consecutive valid measurements were obtained with each device. Both within-device repeatability and between-device agreement were evaluated, while the number of measurement attempts recorded for each analyzer was investigated as an exploratory analysis of feasibility. Results: In total, 56 participants (69.6% male), with a mean age of 70.8 ± 4.3 years and a mean FEV1 of 42.3 ± 13.1% predicted, were enrolled. All analyzers showed excellent within-device repeatability, with ICCs of 0.976 for Hyp'Air, 0.967 for CA2122, and 0.992 for TM2120. Between-device absolute agreement was highest for Hyp'Air versus TM2120 (ICC 0.989), followed by CA2122 versus TM2120 (ICC 0.943) and Hyp'Air versus CA2122 (ICC 0.931). The median number of attempts was 4.5 [3.8-5.0] with Hyp'Air, 3.0 [2.0-3.0] with CA2122, and 2.0 [2.0-3.0] with TM2120 (overall p < 0.001). Conclusions: All three FeNO analyzers demonstrated excellent within-device repeatability in patients with COPD. TM2120 showed closer agreement with Hyp'Air than CA2122, although the limited representation of elevated FeNO values prevents firm conclusions regarding inter-device agreement at higher concentrations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Fractional Exhaled Nitric Oxide Falls Progressively During Treated Exacerbations of COPD and Is Discordant with Blood Eosinophil Counts.Diagnostics (Basel, Switzerland) · 2026
- Fractional exhaled nitric oxide in non-cystic fibrosis bronchiectasis: an exploratory prospective observational study.Monaldi archives for chest disease = Archivio Monaldi per le malattie del torace · 2026
- Chinese Expert Consensus on the Testing Process, Quality Control and Clinical Application of Fractional Exhaled Nitric Oxide (2025).The clinical respiratory journal · 2026
- Diagnostic Accuracy of Tidal Breathing Patterns in Assessing Airflow Obstruction Compared with Spirometry.International journal of chronic obstructive pulmonary disease · 2026
- Short-term longitudinal variation in exhaled nitric oxide fraction predicts future moderate-to-severe exacerbations in COPD patients on triple therapy: a prospective cohort study.ERJ open research · 2026
- Tidal Breathing Exhaled Nitric Oxide Measurement Using A Novel Chemiluminescence Analyzer: Clinical Potential in Lung Disease Diagnosis.Nitric oxide : biology and chemistry · 2026