Analytical evaluation of a urine immunoassay for 2-ethylidene-1,5-dimethyl-3,3-diphenyl pyrrolidine (EDDP) with focus on cutoff optimization.
Valentina Martini, Brian Rossi, Claudia Vignali, Stefania Ortu, Cristiana Stramesi, Luca Morini
Department of Public Health, Experimental and Forensic Medicine, University of Pavia, Pavia, Italy.
Key points
From the paper's conclusion- Overall accuracy remained 99% across all tested cutoff values.
Abstract
Presented is an analytical evaluation of the quantILab Methadone Metabolite (EDDP) immunoassay (Werfen, Italy) performed on the ILab Taurus clinical chemistry analyzer, with assessment of three cutoff concentrations (100, 200, and 300 ng/mL) using liquid chromatography-tandem mass spectrometry (LC-MS/MS) as the confirmatory reference method. A total of 100 human urine samples (50 EDDP-positive and 50 EDDP-negative by routine screening) were analyzed and compared with a confirmatory LC-MS/MS method. Confirmatory LC-MS/MS analysis showed EDDP concentrations ranging from 48.8 to 25,500 ng/mL. At a cutoff of 100 ng/mL, the immunoassay achieved a sensitivity of 100% and a specificity of 98%, with no false-negative results observed. Increasing the cutoff to 200 and 300 ng/mL resulted in a slight decrease in sensitivity (98%) and the occurrence of false-negative results, while specificity reached 100%. Overall accuracy remained 99% across all tested cutoff values. The occurrence of discordant results was limited to specimens with EDDP concentrations close to the evaluated cutoff values, whereas samples with higher EDDP concentrations were consistently classified correctly. These findings demonstrate that cutoff selection significantly influences the diagnostic performance of immunoassay-based screening. In forensic toxicology, where false negatives are particularly critical, a cutoff of 100 ng/mL appears to provide the most appropriate balance between sensitivity and specificity. The evaluated immunoassay represents a reliable tool for high-throughput screening, although confirmatory analysis by LC-MS/MS remains essential.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of Immunoassay Results near Cutoff Values with LC-MS/MS in Individuals with Substance Use Disorder Under Probation.Metabolites · 2026
- Underrecognized Substance Exposure in Pediatric Emergency Departments: Insights from Universal LC-MS/MS Confirmation.Substance use & misuse · 2026
- A-379 Comparative Performance Evaluation of Two Multi-Drug Urine Screening Cups with Fentanyl DetectionClinical Chemistry · 2026
- Detection and quantification of 4F-metomidate and its metabolites, metomidate, in human blood and urine by ultraperformance liquid chromatography with triple quadrupole mass spectrometry (UPLC-MS/MS).Legal medicine (Tokyo, Japan) · 2026
- Tentative identification of a labetalol metabolite in the investigation of a false-positive amphetamine immunoassay result using targeted and untargeted high-resolution mass spectrometry.Clinical chemistry and laboratory medicine · 2026
- Analytical performance and clinical classification value of a novel sandwich immunoassay for total testosterone measurement in low-concentration and hemodialysis populations.Clinical chemistry and laboratory medicine · 2026