Clinical chemistry and laboratory medicine

Tentative identification of a labetalol metabolite in the investigation of a false-positive amphetamine immunoassay result using targeted and untargeted high-resolution mass spectrometry.

Sonali Suresh Karnani-Khemlani, Miguel Ángel Pico-Picos, Jacobo Javier Villacampa-Jiménez, Elene Laburu-Dañobeitia, Darío Cabana-Ramos, Jose Manuel Matey, Luis Manuel Menéndez-Quintanal

Department of Laboratory Medicine, Complejo Hospitalario Universitario de Canarias, San Cristóbal de La Laguna, Spain. · Department of Chemistry and Drugs, National Institute of Toxicology and Forensic Sciences (INTCF), Las Rozas de Madrid, Spain.

Key points

From the paper's conclusion
  • The contrasting immunoassay findings and HRMS results support an assay-dependent analytical interference and illustrate how targeted and untargeted HRMS can progressively refine the investigation of unexpected screening results. However, the tentatively identified labetalol metabolite candidate cannot be established as the cause of the nal von minden response without dedicated cross-reactivity experiments using a purified reference standard.

Abstract

OBJECTIVES: Immunoassays are widely used for urine drug screening in hospital and emergency departments, but assay-dependent cross-reactivity may produce false-positive results. Labetalol and its metabolites have previously been associated with false-positive amphetamine immunoassay findings. We investigated an unexpected positive amphetamine screening result in a 45-year-old pregnant woman receiving labetalol. METHODS: The urine specimen was evaluated using two amphetamine immunoassay methods and subsequently investigated by ultrahigh-performance liquid chromatography-high-resolution tandem mass spectrometuy (UHPLC-HRMS/MS). Targeted HRMS was used to identify known compounds, whereas untargeted HRMS was applied to detect and prioritize unexpected features potentially relevant to the false-positive immunoassay result. Authentic reference standards, when available, were analyzed under identical LC-HRMS conditions to discriminate among relevant isobaric candidates. RESULTS: The urine specimen was positive for amphetamines with the AMP1000 parameter of the nal von minden Drug-Screen® Multi 10TD cassette but negative with the CEDIA amphetamine immunoassay, indicating an assay-dependent false-positive screening result. Targeted HRMS identified labetalol, while untargeted analysis detected a C10H15N feature for which 1-methyl-3-phenylpropylamine(3-amino-1-phenylbutane), a known minor labetalol metabolite, provided the highest spectral-library match. Phentermine and methamphetamine were among the isobaric candidates but were experimentally excluded by analysis of authentic standards. CONCLUSIONS: The contrasting immunoassay findings and HRMS results support an assay-dependent analytical interference and illustrate how targeted and untargeted HRMS can progressively refine the investigation of unexpected screening results. However, the tentatively identified labetalol metabolite candidate cannot be established as the cause of the nal von minden response without dedicated cross-reactivity experiments using a purified reference standard.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers