A-379 Comparative Performance Evaluation of Two Multi-Drug Urine Screening Cups with Fentanyl Detection

Cheryl Mathis, Monica Villanueva, Rachel Law, Taylor Snow, Sara Wyness, Brittany Young

Key points

From the paper's conclusion
  • These findings suggest that further interrogation of the fentanyl cutoff concentration may be needed depending on individual clinical needs.

Abstract

Background Point-of-care urine drug screening (UDS) is a cornerstone tool for monitoring patient medication compliance. Multi-drug lateral flow immunochromatography UDS cups are popular in hospitals and clinics owing to their fast turnaround time and drug class diversity. However, as fentanyl-related overdose and death have increased in the United States, clinical deployment of cups that include fentanyl has become an urgent medical need. Therefore, the objective of this study was to evaluate the accuracy, precision, and stability of two CLIA-waived fentanyl-containing UDS cups (BTNX Rapid-Response™ and Abbott iScreen™) against a reference cup, the Amedica NexScreen™. Methods Represented drugs included amphetamine, secobarbital, buprenorphine, oxazepam, benzoylecgonine, fentanyl, methylenedioxymethamphetamine, methamphetamine, morphine, methadone, oxycodone, phencyclidine, and tetrahydrocannabinol. Cup accuracy was assessed using 25 residual pooled or spiked patient samples, ensuring a minimum of two positive results per drug across the study set. Cup precision was evaluated using blank synthetic urine pools spiked with individual drugs at cutoff, +/-25%, and +/-50% concentrations. Precision samples were run in duplicate over 5 days along with paired positive and negative manufacturer QCs. Accuracy and precision acceptance criteria required at least 90% overall qualitative agreement, with method comparison discrepancies resolved by LC-MS/MS. Finally, cup stability was compared at ambient temperature for both +/-50% concentrations at 5m, 10m, 15m, 30m, 1h, 2h, and 24h by recording changes in the test’s visual readout. Results Both iScreen and Rapid-Response met the overall qualitative accuracy requirements. Considering only LC-MS/MS-confirmed method comparison samples, iScreen produced fewer overall false results (14.3% false positive, 0% false negative) as compared to Rapid-Response (10.7% false positive, 10.7% false negative). Both iScreen and Rapid-Response cups failed to meet overall precision requirements for +/-25% spiked pools but showed improvements at +/-50%. Three analytes showed <90% agreement on iScreen at -50% and 1 failed at +50%; in contrast, Rapid Response precision failed 2 analytes at -50% and 8 failed at +50%. Manufacturer QC analysis produced 100% precision agreement for both cups. The limiting analyte for stability in iScreen is phencyclidine which flipped from positive to negative at the 15m incubation timepoint. Nine of 13 iScreen analytes were stable through 24h at +/-50% concentration levels. The limiting analytes for Rapid-Response are amphetamine, and tetrahydrocannabinol (5m); further, only 3/13 drugs were found to be stable at 24h. Conclusion The iScreen cup demonstrated higher overall accuracy agreement with the NexScreen reference as compared to Rapid-Response. With LC-MS/MS confirmatory testing, iScreen produced only false positives, whereas Rapid-Response generated both false positive and false negatives. This distinction is clinically salient as false positives are reflexed for confirmation, while false negatives are likely missed for follow-up. Although neither cup met precision requirements for all analytes, iScreen exhibited fewer failures and excellent ambient stability. This study found that manufacturer cutoff concentrations were well-supported for 2/13 drugs for Rapid-Response and 9/13 drugs for iScreen. For fentanyl testing, both cups met overall accuracy requirements yet failed by precision for +50%. These findings suggest that further interrogation of the fentanyl cutoff concentration may be needed depending on individual clinical needs.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers