Correlation of NEC to Image SNR and Lesion Detectability in clinical TOF PET.
Majid Kazemi Kozani, Srilalan Krishnamoorthy, Joel S Karp, Suleman Surti
Department of Radiology of the University of Pennsylvania, Philadelphia, PA 19104.
Key points
From the paper's conclusion- The results indicate that while image SNR scales approximately linearly with NEC, this relationship becomes substantially more consistent when NECTOF is used, demonstrating that NECTOF is a better predictor of global image SNR under iterative reconstruction than conventional NEC.
Abstract
In positron emission tomography (PET) imaging, Noise Equivalent Counts (NEC) serve as a key indicator of overall system performance. Previous studies have established the relationship between NEC and signal-to-noise (SNR) for both non-time-of-flight (non-TOF) and TOF PET scanners employing filtered back-projection (FBP) algorithms. This study aims to explore the correlation between NEC and image SNR, as well as task-specific metrics, using iterative reconstruction with comprehensive TOF PET system modeling. The phantom, 35 cm in diameter to represent a patient with high BMI, featured twelve 1-cm diameter spheres placed in an off centered slice, and data were acquired across a wide range of activity levels using the PennPET Explorer. Multiple scans were generated via bootstrapping and reconstructed using list-mode TOF-ordered subset expectation maximization (TOF-OSEM) algorithms to obtain mean and standard deviation images. Image SNR was quantified as the ratio of mean values within a central volume-of-interest (VOI) across both images. Utilizing a generalized scan-statistic model, the area under the localization receiver operating characteristic (ALROC) was calculated for the spheres placed at the offset slice. NEC was computed using both conventional and TOF-adapted formulas from the original data. The results indicate that while image SNR scales approximately linearly with NEC, this relationship becomes substantially more consistent when NECTOF is used, demonstrating that NECTOF is a better predictor of global image SNR under iterative reconstruction than conventional NEC. Furthermore, our results show that although ALROC is proportional to SNR, it does not adequately represent the change in ALROC as TOF resolution improves.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Source-Traced PET/CT Scanner Performance Database (2004-2024): The Sphere-Size Dependence of Time-of-Flight Contrast Recovery.Diagnostics (Basel, Switzerland) · 2026
- Toward an optimized PET reconstruction protocol for spinal cord: superiority of OSEM with TOF and PSF.EJNMMI physics · 2026
- Attenuation Pre-Correction in Iterative PET Image Reconstruction: What is the Statistical Cost?IEEE transactions on radiation and plasma medical sciences · 2026
- Attenuation corrected activity reconstruction for listmode TOF-PET data.ArXiv · 2026
- Iterative Reconstruction Performance on the NeuroEXPLORER: Evaluation with Static and Dynamic Human Imaging.IEEE transactions on radiation and plasma medical sciences · 2026
- Per-Bundle, Pre-Reconstruction Inversion for Temporal CT: Flux-Compensation Design and Multi-Patient GeneralizationarXiv · 2026