A retrospective multi-center validation study of tomography-based formulas for phakic intraocular lens sizing.
William Goes, Isabelle Saelens, Frank Goes, Elisa Mulier, Heleen Delbeke
Department of ophthalmology, University Hospitals Leuven, Leuven, Belgium.| UZ Leuven. · Department of ophthalmology, University Hospitals Leuven, Leuven, Belgium. KU Leuven, Biomedical Sciences Group, Department of Neurosciences, Research Group Ophthalmology, Leuven, Belgium.| UZ Leuven.
Key points
From the paper's conclusion- In this study the accuracy of the LASSO-formulas varies across the different models and according to the postoperative measurement method, with the LASSO-OCT formula outperforming the LASSO-Full when vault is measured with AS-OCT. Other OCT based formulas, including NK V3 and Russo, demonstrated high accuracy and reliability, supporting the safety of OCT-based ICL sizing.
Abstract
PURPOSE: To validate the Least Absolute Shrinkage and Selection Operator (LASSO) formulas for phakic intraocular lens sizing, published by Rocamora et al. in 2023, in a larger and multicenter cohort. SETTING: University hospital and private clinical practice. DESIGN: Multicenter retrospective observational study in Belgium. METHODS: Patients who underwent implantable Collamer lens (ICL) surgery between September 2018 and June 2024 were included. The mean absolute error (MAE) was used as primary endpoint to compare LASSO-OCT, LASSO-Biometry, LASSO-Full, Luo, Russo, Kim, Nakamura (NK) V1, NK V2 and NK V3 formulas. Secondary endpoints included the maximum absolute error and the percentage of eyes within error categories of <100, <200, <300, <500, <750 µm. RESULTS: A total of 437 eyes of 226 patients were included. The pooled MAE of the LASSO-OCT formula (155.6 ±123.0 µm) was lower than that of the LASSO-Biometry (193.3±140.3 µm) and LASSO-Full (247.2 ±174.7 µm) formulas. The Russo formula showed the lowest MAE (142.3 ±112.4 µm) followed by the NK V3 (149.3 ±121.0 µm). The MAE was significantly higher for the LASSO-Full, Luo, NK V1 and V2 formulas when compared with the LASSO-OCT formula. At the center where postoperative vault was measured using Scheimpflug imaging, the LASSO-Full formula achieved a lower MAE than the LASSO-OCT formula, although this difference was not statistically significant. CONCLUSION: In this study the accuracy of the LASSO-formulas varies across the different models and according to the postoperative measurement method, with the LASSO-OCT formula outperforming the LASSO-Full when vault is measured with AS-OCT. Other OCT based formulas, including NK V3 and Russo, demonstrated high accuracy and reliability, supporting the safety of OCT-based ICL sizing.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Biometry-Guided IOL Formula Selection in Post-Laser Vision Correction Eyes.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Comparison of the Accuracy of the Web-Based ESCRS IOL Power Calculator Formulas.Seminars in ophthalmology · 2026
- Accuracy of NK and KS Formulas in Predicting Postoperative Vault at One Month After Implantable Collamer Lens Implantation: A Retrospective Study of 327 Eyes.Clinical ophthalmology (Auckland, N.Z.) · 2026
- Comparison of the Moussa Intraocular Lens Formula to contemporary formulas in a multicenter study.Journal of cataract and refractive surgery · 2026
- Biometry-Driven Selection of IOL Power Calculation Formulas Using the ESCRS IOL Calculator.Journal of cataract and refractive surgery · 2026
- Comparison of IOL power calculation formulas in short eyes using ESCRS calculator and eyetemis: Influence of anterior segment biometry.PloS one · 2026