Accuracy of NK and KS Formulas in Predicting Postoperative Vault at One Month After Implantable Collamer Lens Implantation: A Retrospective Study of 327 Eyes.
Zixiao Liu, Xinhua Liu, Xue Yao, Jiantao Wang, Jing Zhang
Shenzhen Eye Hospital; Shenzhen Eye Medical Center, Southern Medical University, Shenzhen, People's Republic of China.
Key points
From the paper's conclusion- The KS formula showed better agreement with achieved vault and more frequently recommended the implanted ICL size than the NK formula. However, both formulas showed variable accuracy across ICL sizes, highlighting the need for further refinement to optimize ICL size selection. Because vault was assessed at one month, these findings should be regarded as short-term observations.
Abstract
PURPOSE: To evaluate the agreement and accuracy of the Nakamura-Kamiya (NK) and Kimura-Sato (KS) formulas in predicting postoperative vault at one month after implantable collamer lens (ICL) implantation, and to assess the agreement between the ICL sizes recommended by these formulas and the sizes actually implanted. METHODS: This retrospective study included 327 eyes of 168 patients who underwent ICL V4c implantation. Predicted vaults from the NK and KS formulas were compared with achieved vaults at one month. Linear mixed-effects models and generalized estimating equations were used to account for inter-eye correlation, with a one-eye sensitivity analysis. Agreement was assessed using Bland-Altman analysis, the intraclass correlation coefficient, and Cohen's kappa. Prediction accuracy was assessed using mean absolute error(MAE) and root mean square error(RMSE). RESULTS: The actual ideal vault rate was 63.61%. The NK formula overestimated vault (median 664 µm vs 574 µm; P < 0.001), whereas the KS formula slightly underestimated it (562 µm vs 574 µm; P = 0.040). The KS formula showed better consistency with actual vault than the NK formula (R 2 = 0.264 vs 0.147) and narrower 95% limits of agreement on Bland-Altman analysis.In mixed-effects models, the NK slope was 0.385 (95% CI, 0.238-0.532) and the KS slope was 1.117 (95% CI, 0.856-1.378; both P < 0.001). MAE was 225.91 µm (NK) and 184.18 µm (KS); RMSE was 291.35 µm (NK) and 246.86 µm(KS), respectively. The NK formula recommended the implanted size in 60.24% of eyes and the KS formula in 83.79% (kappa 0.283 and 0.729). Ultrasound biomicroscopy (UBM) measured anterior chamber depth(ACD),crystalline lens rise(CLR), horizontal and vertical sulcus-to-sulcus (STS), and age were independently associated with achieved vault (All P < 0.05). CONCLUSION: The KS formula showed better agreement with achieved vault and more frequently recommended the implanted ICL size than the NK formula. However, both formulas showed variable accuracy across ICL sizes, highlighting the need for further refinement to optimize ICL size selection. Because vault was assessed at one month, these findings should be regarded as short-term observations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Research on a New Model for Predicting Vault After Implantable Collamer Lens (ICL) Based on Structural Parameters of the Posterior Chamber.Journal of ophthalmology · 2026
- Development and Clinical Evaluation of an Anterior Chamber Angle-Based Lens Sizing Method for Implantable Collamer Lens Surgery.Clinical ophthalmology (Auckland, N.Z.) · 2026
- ICL Sizing and Vault Prediction: A Systematic Review of Mathematical Models and Calculation Formulas.Seminars in ophthalmology · 2026
- A retrospective multi-center validation study of tomography-based formulas for phakic intraocular lens sizing.Journal of cataract and refractive surgery · 2026
- Retrospective Validation of an AI Based Tool for Vault Prediction After Implantable Collamer Lens Surgery for High Myopia.Clinical ophthalmology (Auckland, N.Z.) · 2026
- Effect of implantable collamer lens on biometric measurements and intraocular lens power calculation formulas.Acta ophthalmologica · 2026