Detection of Endothelial Plaque in Microbial Keratitis Using Anterior Segment Optical Coherence Tomography: Findings From the METRIC Study.
Folahan Ibukun, Kamini Reddy, Subeesh Kuyyadiyil, Elesh Jain, Gautam Parmar, Nakul S Shekhawat
Wilmer Eye Institute, Johns Hopkins University School of Medicine, Baltimore, MD; and. · SNC Chitrakoot, Chitrakoot, Madhya Pradesh, India.
Key points
From the paper's conclusion- SLE substantially underdetects endothelial plaques compared with ASOCT, which enables reproducible detection with excellent agreement. Outcome-based validation of the significance of this increased detection is warranted.
Abstract
PURPOSE: To evaluate intragrader and intergrader concordance of anterior segment optical coherence tomography (ASOCT) for detecting endothelial plaque in microbial keratitis, and compare ASOCT detection versus slit lamp examination (SLE). METHODS: Diagnostic concordance study of 150 eyes with microbiologically confirmed bacterial or fungal keratitis at a tertiary eye hospital. Two masked ophthalmologist graders independently evaluated ASOCT images for 2 morphologically distinct endothelial plaque subtypes (round and flat). We assessed intragrader and intergrader concordance and diagnostic agreement between ASOCT and SLE using kappa, sensitivity, and specificity. Logistic regression assessed factors associated with disagreement between ASOCT and SLE after adjustment for covariates. RESULTS: ASOCT showed near perfect intergrader agreement for detecting round endothelial plaques (kappa 0.88), flat endothelial plaques (kappa 0.84), and either plaque (kappa 0.88). Intragrader agreement was substantial to near perfect for both graders (kappa 0.70-0.86). SLE identified endothelial plaque in 6.0% eyes, whereas ASOCT detected round plaques in 32.7%, flat plaques in 43.3%, and either plaque in 55.3% of eyes. Using ASOCT as reference, slit lamp sensitivity was 16.3% for round plaques, 6.2% for flat plaques, and 9.6% for either, with specificity exceeding 94%. Poor visual acuity (logarithm of the minimum angle of resolution ≥1.0) was associated with disagreement for round plaques (adjusted odds ratio [OR] 5.04), flat plaques (adjusted OR 3.63), and either morphology (adjusted OR 3.98). Bacterial infection was associated with disagreement for any plaque (adjusted OR 4.56). CONCLUSIONS: SLE substantially underdetects endothelial plaques compared with ASOCT, which enables reproducible detection with excellent agreement. Outcome-based validation of the significance of this increased detection is warranted.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Agreement between Sirius Scheimpflug-Placido tomography and Anterion swept-source OCT in keratoconic eyes stratified by prior corneal cross-linking: a retrospective cross-sectional agreement study.International ophthalmology · 2026
- Comparison of spectral-domain optical coherence tomography biometry measurements with partial coherence interferometry optical biometryBMC Ophthalmology · 2026
- Establishing a precision reference benchmark for corneal epithelial aberration measurements using a new hybrid device across keratoconus stages.Eye and vision (London, England) · 2026
- Extending a Standard SD-OCT Device to Ocular Biometry with Additional Software: Agreement and Clinical Applicability Compared with Ultrasound and Optical Biometry.Journal of clinical medicine · 2026
- Automated quantification of central corneal epithelial thickness for the diagnosis of limbal stem cell deficiency.PloS one · 2026
- Optical Biometry Errors in Eyes with Iris-Claw Phakic Intraocular Lenses.Clinical ophthalmology (Auckland, N.Z.) · 2026