Optical Biometry Errors in Eyes with Iris-Claw Phakic Intraocular Lenses.
Wei-Lun Huang, Shao-Wei Huang, Ya-Ting Chiu, Huai-Wen Chang, Hsiao-Ling Shen, Wei-Jen Huang, Fung-Rong Hu
Department of Ophthalmology, National Taiwan University Hospital Hsinchu Branch, Hsinchu, Taiwan. · Department of Education, National Taiwan University Hospital, College of Medicine, National Taiwan University, Taipei, Taiwan.
Key points
From the paper's conclusion- In eyes with iris-claw pIOLs, automated optical biometers may misidentify ocular structures and underestimate ACD, potentially leading to clinically meaningful differences in hypothetical IOL power selection. Manual review and correction of Eyestar B-scan images restored ACD measurements to close agreement with Pentacam values. These findings highlight the importance of careful inspection of raw biometry images in such eyes, particularly when using formulas that incorporate measured ACD.
Abstract
PURPOSE: To evaluate the effect of iris-claw phakic intraocular lenses (pIOLs) on anterior chamber depth (ACD) measurement by optical biometers and the potential impact of ACD errors on hypothetical intraocular lens (IOL) power selection for cataract surgery. PATIENTS AND METHODS: Seventy-six eyes of 38 patients with iris-claw pIOLs (Artiflex) underwent biometry using Pentacam (Scheimpflug), Lenstar LS-900 (optical low-coherence reflectometry), and Eyestar 900 (swept-source OCT). The failure and misidentification rates of ACD measurement were recorded. Eyestar B-scans were manually reviewed and corrected when segmentation errors were identified. Hypothetical IOL power was calculated using Barrett Universal II and SRK/T formulas. RESULTS: Pentacam successfully measured ACD in all 76 eyes. Lenstar LS-900 angd Eyestar 900 failed to provide ACD measurements in 4 eyes (5.3%) and 14 eyes (18.4%), respectively. Using a predefined threshold of >10% deviation from the Pentacam value, inaccurate ACD measurements were identified in 30 eyes (39.5%) with Lenstar and in 17 eyes (22.4%) with Eyestar. On Eyestar, 20 eyes (26.3%) showed misidentification of the anterior pIOL surface as the anterior capsule of the crystalline lens, and 11 eyes (14.5%) were interpreted as aphakic. After manual correction, Eyestar ACD values closely matched Pentacam measurements, with a mean difference of 0.04 ± 0.03 mm. Correction of erroneous ACD measurements changed Barrett Universal II-based hypothetical IOL power selection by at least 0.50 D in 7 of 17 affected eyes (41.2%). CONCLUSION: In eyes with iris-claw pIOLs, automated optical biometers may misidentify ocular structures and underestimate ACD, potentially leading to clinically meaningful differences in hypothetical IOL power selection. Manual review and correction of Eyestar B-scan images restored ACD measurements to close agreement with Pentacam values. These findings highlight the importance of careful inspection of raw biometry images in such eyes, particularly when using formulas that incorporate measured ACD.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy of intraocular lens power calculation in pediatric aphakia with the Artisan iris-claw lens.Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie · 2026
- Use of the internal astigmatism axis to determine the toric intraocular lens axis: Comparison between C-loop and plate-haptic lenses.Arquivos brasileiros de oftalmologia · 2026
- The Repeatability and Agreement of Two SS-OCT Optical Biometers in Adult Eyes Undergoing Cataract Surgery.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Hydrophobic Acrylic Trifocal IOL in Post-Laser Refractive Surgery Eyes: Visual Performance, Predictability, and Patient-Reported Outcomes.Journal of ophthalmology · 2026
- Effect of implantable collamer lens on biometric measurements and intraocular lens power calculation formulas.Acta ophthalmologica · 2026
- Concordance of Cataract Biometry Measurements Obtained via Swept-Source Optical Coherence Tomography Versus Optical Biometry.Bioengineering (Basel, Switzerland) · 2026