Flanged Intraluminal Suture Occlusion of an Ahmed Glaucoma Valve Tube During Suprachoroidal Hemorrhage Drainage: A Surgical Technique.
Semir Yarımada, Sinem Karabulut, Pınar Beyazgül, Müjdat Karabulut
Department of Ophthalmology, Tepecik Training and Research Hospital, İzmir, Turkiye. · Department of Ophthalmology, İzmir University of Economics Faculty of Medicine, İzmir, Turkiye.
Key points
From the paper's conclusion- A flanged intraluminal polypropylene suture plug may limit iatrogenic AC pressure transmission to an AGV valve during concurrent SCH drainage and can be left in place indefinitely when an alternative outflow pathway maintains IOP control; whether the valve mechanism itself was protected from pressure-related injury was not directly tested in this case.
Abstract
PURPOSE: To describe a technique for temporarily occluding an Ahmed glaucoma valve (AGV) tube with a flanged, intraluminal polypropylene suture plug to limit transmission of anterior chamber (AC) pressure to the valve membrane during trocar-assisted drainage of a delayed suprachoroidal hemorrhage (SCH). METHODS: Case report of a 74-year-old man with bilateral advanced glaucoma and bilaterally nonfunctional filtering blebs who underwent AGV implantation in the left eye. After a coughing-induced hypotony episode on postoperative day 3, a kissing, four-quadrant SCH was diagnosed by B-scan ultrasonography and drained under general anesthesia 2 weeks later via small-gauge trocars, after first occluding the AGV tube with an extraocularly flanged 3-0 polypropylene suture introduced through a dedicated corneal side-port. RESULTS: The previously nonfunctional trabeculectomy bleb was observed to reactivate coincident with the hypotony episode and subsequently sustained intraocular pressure (IOP) control without medication (approximately 11 mmHg at 2 months) once the AGV tube was permanently occluded. Best-corrected visual acuity (BCVA) improved from 20/200 (0.1 decimal) at 2 weeks to 20/70 (0.3 decimal) at 2 months postoperatively, with no recurrence of choroidal effusion or SCH and no suture-related complications. CONCLUSION: A flanged intraluminal polypropylene suture plug may limit iatrogenic AC pressure transmission to an AGV valve during concurrent SCH drainage and can be left in place indefinitely when an alternative outflow pathway maintains IOP control; whether the valve mechanism itself was protected from pressure-related injury was not directly tested in this case.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The FIRE technique - A tube ligation alternative.Indian journal of ophthalmology · 2026
- Novel Dual-Device Strategy: Failed PRESERFLO MicroShunt Connected to an Ahmed ClearPath - Case Report.Case reports in ophthalmology · 2026
- Management of Ahmed Glaucoma Valve Plate Exposure by Plate Removal and Tube Retention: A Case Report.International medical case reports journal · 2026
- Six-Month Outcomes of Ahmed Glaucoma Valve Implantation Using a Patch Graft-Free Scleral Tunnel Technique in Eyes with Refractory Glaucoma: A prospective, non-comparative interventional study.La Clinica terapeutica · 2026
- Recombinant Tissue Plasminogen Activator-assisted Early Drainage of Suprachoroidal Hemorrhage.Journal of current glaucoma practice · 2026
- Conjunctiva-scleral thickness over the Ahmed glaucoma valve tube in eyes at clinically suspected risk of tube exposure: Anterior segment optical coherence tomography findings.Arquivos brasileiros de oftalmologia · 2026