Indian journal of ophthalmology

The FIRE technique - A tube ligation alternative.

Roya Garakani, Jamie Beckman, Iqbal Ike K Ahmed

Department of Ophthalmology, John A. Moran Eye Center, University of Utah, Salt Lake City, UT, USA. · Department of Ophthalmology, Prism Eye Institute, Mississauga, Canada.

Key points

From the paper's conclusion
  • Here, we propose an alternative approach, the Flanged Intraluminal Ripcord Element (FIRE) technique, that eliminates the need for a ligature suture and avoids the unpredictability associated with ligature autolysis.

Abstract

Nonvalved glaucoma drainage devices (GDDs) carry a risk of early hypotony due to unregulated flow prior to capsule formation. Temporary flow restriction occurs through methods such as absorbable ligature sutures and intraluminal ripcords. Here, we propose an alternative approach, the Flanged Intraluminal Ripcord Element (FIRE) technique, that eliminates the need for a ligature suture and avoids the unpredictability associated with ligature autolysis. A patient case is discussed that demonstrated adequate prevention of hypotony and successful intraocular pressure reduction upon ripcord removal. The FIRE technique provides predictable flow control in the early postoperative period of nonvalved GDDs.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers