The FIRE technique - A tube ligation alternative.
Roya Garakani, Jamie Beckman, Iqbal Ike K Ahmed
Department of Ophthalmology, John A. Moran Eye Center, University of Utah, Salt Lake City, UT, USA. · Department of Ophthalmology, Prism Eye Institute, Mississauga, Canada.
Key points
From the paper's conclusion- Here, we propose an alternative approach, the Flanged Intraluminal Ripcord Element (FIRE) technique, that eliminates the need for a ligature suture and avoids the unpredictability associated with ligature autolysis.
Abstract
Nonvalved glaucoma drainage devices (GDDs) carry a risk of early hypotony due to unregulated flow prior to capsule formation. Temporary flow restriction occurs through methods such as absorbable ligature sutures and intraluminal ripcords. Here, we propose an alternative approach, the Flanged Intraluminal Ripcord Element (FIRE) technique, that eliminates the need for a ligature suture and avoids the unpredictability associated with ligature autolysis. A patient case is discussed that demonstrated adequate prevention of hypotony and successful intraocular pressure reduction upon ripcord removal. The FIRE technique provides predictable flow control in the early postoperative period of nonvalved GDDs.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative In Vitro Evaluation of Tube Resistance in Glaucoma Drainage Devices with Intraluminal Polypropylene Threads.Bioengineering (Basel, Switzerland) · 2026
- Flanged Intraluminal Suture Occlusion of an Ahmed Glaucoma Valve Tube During Suprachoroidal Hemorrhage Drainage: A Surgical Technique.Retinal cases & brief reports · 2026
- Efficacy and safety of a novel glaucoma drainage tube with a porous outer structure (GDTI-50H) in rabbit eyes.Journal of materials science. Materials in medicine · 2026
- Novel surgical approach to correct late-hypotony in PreserFlo MicroShunt.The British journal of ophthalmology · 2026
- Nd: YAG laser rescue of sulcus-placed glaucoma drainage device tube occlusion: a two-case series.BMC ophthalmology · 2026
- Optimizing irrigation and improving surgical operability in UBE: Technical note and preliminary results of a new guiding device.Medicine · 2026