Empirical and semi-empirical methods to determine the kV exponent (n value) for diagnostic X-ray machines.
Wellen Rukundo, Natharius Nimbashabira, Samuel Wattisa Maberi, Christine Edna Emong
Department of Regulatory Affairs, Atomic Energy Council, Plot 40, Bukoto Street, P.O. Box 7044, Kampala, Uganda.
Key points
From the paper's conclusion- Therefore, the study introduces a semi-empirical method for estimating the exponent ($n$ value) that incorporates measured half-value layer variation to account for effective beam hardening, which is neglected in conventional empirical models.
Abstract
The air kerma, and beam quality measurements were performed on 43 diagnostic X-ray machines (25 fixed and 18 mobile) that met acceptance criteria for tube potential accuracy, half-value layer, and tube output. Measurements were conducted at tube potentials from 40 kV to 120 kV using a calibrated solid-state dosimetry system at a fixed focus-detector distance of 1 m, normalized output (mGy/mAs) and half-value layer were determined at each setting of the tube potential. Two methods (method 1 - empirical and method 2 - semi-empirical) were used to derive the exponent value. In method 1, a power-law fit of tube output versus tube potential yielded an average exponent of $n$ ≈ 2.334 ± 0.118 based on the group of X-ray machines with high-power frequency generators, consistent with values reported for diagnostic systems with standard filtration [1, 2]. In method 2, the $n$ was derived as a function of tube potential by incorporating the measured variation of half-value layer with tube potential, explicitly accounting for effective filtration. This approach showed that $n$ decreases from ~2.69 at 40 kV to 2.09 at 120 kV, demonstrating greater output sensitivity to tube potential changes at lower tube potentials. Therefore, the study introduces a semi-empirical method for estimating the exponent ($n$ value) that incorporates measured half-value layer variation to account for effective beam hardening, which is neglected in conventional empirical models. This provides a practical tool for characterizing the dependence of X-ray tube output on tube protocol optimization and equipment performance assessment, applicable for quality control in diagnostic X-ray machines.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Experimental and Monte Carlo validation of DIN 6809-4:2020 kilovoltage X-ray beam qualities for secondary standard dosimetry applications.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Calibration of X-ray multimeters for radiography.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Calibration of X-ray multimeters for half-value layer: measurement methods and uncertainty evaluationMetrologia · 2026
- Experimental Comparison of Half-Value Layer Measurements Using a Solid-State X-ray Meter and an Ionization Chamber in General Radiography.Cureus · 2026
- DNA double-strand break yield and radiation quality of diagnostic X-rays from 40 to 120 kV: a scale-resolved microdosimetric and track-structure studybioRxiv · 2026
- Accuracy of reference air kerma in estimating incident air kerma at skin in interventional neuroradiology: Effect of X-ray field size and alternative PERP corrections.Journal of applied clinical medical physics · 2026