Calibration of X-ray multimeters for radiography.
Paula Toroi, Viivi Valkama, Touko Kaasalainen, Niko Kiiskinen, Mika Kortesniemi, Panu Pousi, Jaroslav Šolc, Aino Tietäväinen
STUK - Radiation and Nuclear Safety Authority, Vantaa, Finland; HUS Diagnostic Center, Radiology, Helsinki University and Helsinki University Hospital, Helsinki, Finland. Electronic address: paula.toroi@stuk.fi. · STUK - Radiation and Nuclear Safety Authority, Vantaa, Finland.
Key points
From the paper's conclusion- XMMs can provide reliable measurements across clinically relevant radiation qualities, including copper‑filtered beams, when radiation‑quality‑specific calibration coefficients are applied. Interpolation based on a single radiation quality parameter is insufficient; at least two parameters are required.
Abstract
INTRODUCTION: Accurate determination of air kerma, tube‑voltage‑related quantities (TV), and half‑value layer (HVL) is essential for reliable dosimetry in X‑ray imaging. X‑ray multimeters (XMMs) are widely used for these measurements; however, their performance depends on spectral characteristics and measurement conditions. This study evaluates the consistency and clinical applicability of calibration coefficients derived using industrial and clinical X‑ray systems across a range of clinically relevant radiation qualities. METHODS: Eight XMMs from four manufacturers were calibrated for air kerma, TV, and HVL using one laboratory and three clinical systems. Reference values were established using traceable ionization chambers, high voltage dividers, and spectrometry. Calibration coefficients were assessed for standard aluminium filtered radiation qualities and for an extended range including 0-0.9 mm of copper filtration. Energy dependence, X‑ray system dependence, and the impact of applying calibration coefficients were evaluated. RESULTS: For air kerma, calibration coefficients derived using industrial and clinical systems were highly consistent. Energy dependence variations were within 5% and could be reduced to below 1% by applying radiation‑quality‑specific calibration coefficients. TV measurements showed the strongest dependence on the X‑ray system; energy dependence variations were generally below 4% and could be improved only using clinical‑system‑based calibrations. For HVL measurements, energy dependence variations were typically within 6% and were reduced below 1% by calibration, except for two XMMs exhibiting system‑dependent behaviour. CONCLUSIONS: XMMs can provide reliable measurements across clinically relevant radiation qualities, including copper‑filtered beams, when radiation‑quality‑specific calibration coefficients are applied. Interpolation based on a single radiation quality parameter is insufficient; at least two parameters are required.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Calibration of X-ray multimeters for half-value layer: measurement methods and uncertainty evaluationMetrologia · 2026
- Performance of air kerma - area product meters based on long-term calibration data.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Experimental and Monte Carlo validation of DIN 6809-4:2020 kilovoltage X-ray beam qualities for secondary standard dosimetry applications.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- Empirical and semi-empirical methods to determine the kV exponent (n value) for diagnostic X-ray machines.Radiation protection dosimetry · 2026
- Impact of a common user-derived electrometer calibration coefficient on absorbed dose-to-water determination: a multicenter uncertainty studyJournal of Radioanalytical and Nuclear Chemistry · 2026
- Experimental Comparison of Half-Value Layer Measurements Using a Solid-State X-ray Meter and an Ionization Chamber in General Radiography.Cureus · 2026