Experimental and Monte Carlo validation of DIN 6809-4:2020 kilovoltage X-ray beam qualities for secondary standard dosimetry applications.

Kosar Farajpooredari, Nematollah Heidarloo, Seyed Mahmoud Reza Aghamiri, Bahador Bahadorzadeh

Department of Medical Radiation Engineering, Shahid Beheshti University, P.O. Box 19839-69411, Tehran, Iran. · Department of Medical Radiation Engineering, Shahid Beheshti University, P.O. Box 19839-69411, Tehran, Iran. Electronic address: n_heidarloo@sbu.ac.ir.

Key points

From the paper's conclusion
  • The results demonstrate that the combined experimental-Monte Carlo methodology provides a practical approach for independent validation of DIN-compliant kilovoltage beam qualities and may support improved dosimetric traceability in secondary standard dosimetry laboratories.

Abstract

Kilovoltage radiotherapy (kV-RT) is a well-established modality for treating superficial lesions, where accurate dosimetry is essential for treatment efficacy and patient safety. This study presents an experimental and Monte Carlo-based framework for the validation of DIN 6809-4:2020 standard X-ray beam qualities at Secondary Standard Dosimetry application. A kilovoltage MXR-320/26 X-ray tube was employed. The tube's inherent filtration was characterized via measurement of the first half-value layer (HVL1) at 60 kV per ISO 4037-1, yielding an equivalent filtration of 0.13 mm Al. Standard beam qualities for the soft X-ray (TW, 30-100 kV) and hard X-ray (TH, 120-300 kV) series were established using calibrated PTW ionization chambers (23344 and 30013). Given the unavailability of direct X-ray spectrometry, an MCNPX Monte Carlo model was developed for independent spectral validation. Experimental HVL1 measurements showed good agreement with DIN reference values, with mean absolute deviations of 1.51% ± 0.77% (TW) and 1.45% ± 0.93% (TH), remaining within the ±5% ISO tolerance. MC simulations reproduced the spectral characteristic with mean absolute deviations of 1.50% ± 0.75% for HVL1 and 0.59% ± 0.38% for mean photon energy. A GUM-compliant uncertainty analysis yielded expanded uncertainties (k = 2) of 1.3% for Group 1 (TW-30 to TW-70) and 2.3% for Group 2 (TW-100 and TH-120 to TH-300), consistent with international SSDL performance criteria. Short-term repeatability was better than 0.10%, with long-term stability remaining within ±0.50% over six months. The results demonstrate that the combined experimental-Monte Carlo methodology provides a practical approach for independent validation of DIN-compliant kilovoltage beam qualities and may support improved dosimetric traceability in secondary standard dosimetry laboratories.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers