Experimental and Monte Carlo validation of DIN 6809-4:2020 kilovoltage X-ray beam qualities for secondary standard dosimetry applications.
Kosar Farajpooredari, Nematollah Heidarloo, Seyed Mahmoud Reza Aghamiri, Bahador Bahadorzadeh
Department of Medical Radiation Engineering, Shahid Beheshti University, P.O. Box 19839-69411, Tehran, Iran. · Department of Medical Radiation Engineering, Shahid Beheshti University, P.O. Box 19839-69411, Tehran, Iran. Electronic address: n_heidarloo@sbu.ac.ir.
Key points
From the paper's conclusion- The results demonstrate that the combined experimental-Monte Carlo methodology provides a practical approach for independent validation of DIN-compliant kilovoltage beam qualities and may support improved dosimetric traceability in secondary standard dosimetry laboratories.
Abstract
Kilovoltage radiotherapy (kV-RT) is a well-established modality for treating superficial lesions, where accurate dosimetry is essential for treatment efficacy and patient safety. This study presents an experimental and Monte Carlo-based framework for the validation of DIN 6809-4:2020 standard X-ray beam qualities at Secondary Standard Dosimetry application. A kilovoltage MXR-320/26 X-ray tube was employed. The tube's inherent filtration was characterized via measurement of the first half-value layer (HVL1) at 60 kV per ISO 4037-1, yielding an equivalent filtration of 0.13 mm Al. Standard beam qualities for the soft X-ray (TW, 30-100 kV) and hard X-ray (TH, 120-300 kV) series were established using calibrated PTW ionization chambers (23344 and 30013). Given the unavailability of direct X-ray spectrometry, an MCNPX Monte Carlo model was developed for independent spectral validation. Experimental HVL1 measurements showed good agreement with DIN reference values, with mean absolute deviations of 1.51% ± 0.77% (TW) and 1.45% ± 0.93% (TH), remaining within the ±5% ISO tolerance. MC simulations reproduced the spectral characteristic with mean absolute deviations of 1.50% ± 0.75% for HVL1 and 0.59% ± 0.38% for mean photon energy. A GUM-compliant uncertainty analysis yielded expanded uncertainties (k = 2) of 1.3% for Group 1 (TW-30 to TW-70) and 2.3% for Group 2 (TW-100 and TH-120 to TH-300), consistent with international SSDL performance criteria. Short-term repeatability was better than 0.10%, with long-term stability remaining within ±0.50% over six months. The results demonstrate that the combined experimental-Monte Carlo methodology provides a practical approach for independent validation of DIN-compliant kilovoltage beam qualities and may support improved dosimetric traceability in secondary standard dosimetry laboratories.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Calibration of X-ray multimeters for half-value layer: measurement methods and uncertainty evaluationMetrologia · 2026
- Empirical and semi-empirical methods to determine the kV exponent (n value) for diagnostic X-ray machines.Radiation protection dosimetry · 2026
- Calibration of X-ray multimeters for radiography.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- Measurement-based dosimetric comparison of kilovoltage and megavoltage X-ray systems for static-field 3D conformal radiotherapy in companion animals.Journal of veterinary science · 2026
- AAPM Medical Physics Practice Guideline 18.a: Superficial radiation therapy using low-energy kilovoltage x-rays.Journal of applied clinical medical physics · 2026
- DNA double-strand break yield and radiation quality of diagnostic X-rays from 40 to 120 kV: a scale-resolved microdosimetric and track-structure studybioRxiv · 2026