Calculation and correction of the eye-lens dosemeter response at non-calibration photon energies.
Samaneh Baradaran, Amir Moslehi, Aram Rostami, Mehran Taheri
Nuclear Science and Technology Research Institute, North Karegar St., Tehran, P.O. Box: 11365-3486, Iran. · Radiation Applications Research School, Nuclear Science and Technology Research Institute, North Karegar St., Tehran, P.O. Box: 11365-3486, Iran.
Key points
From the paper's conclusion- The results indicated that the response at energies below 200 keV deviates by $\sim$1.85 times compared with that at 662 keV.
Abstract
In this article, the calibration curve of GR-200 TLDs (Thermo-Luminescent Dosemeters) as eye-lens dosemeters was experimentally generated in a standard $^{137}$Cs field. Subsequently, the energy and angular responses of TLDs positioned on the ISO cylindrical head phantom were calculated using Geant4 simulation toolkit. The results indicated that the response at energies below 200 keV deviates by $\sim$1.85 times compared with that at 662 keV. Furthermore, the responses to 70 kVp and 120 kVp X-rays used in angiography were 1.65 and 1.61 times greater, respectively. The responses to these X-rays were also found to be independent of the irradiation angle between 0$^{\circ }$ and 75$^{\circ }$, with a maximum deviation of 28%. Finally, a correction method based on the relative energy response of GR-200 TLDs was proposed to adjust their response at photon energies or spectra different from the calibration field, thereby improving the accuracy of H$_{p}$(3) assessment.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of the Measurement Uncertainty of Thermoluminescence Dosimeters (TLD-100 and GR-200) in Clinical Radiotherapy Energies.Journal of biomedical physics & engineering · 2026
- Validation of a 3D-printed anthropomorphic eye phantom for quality assurance in external beam radiotherapy using TLD and radiochromic film.Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine · 2026
- TL/OSL emission spectrometry of alumina substrates of electronic components in mobile phones: potential of the red TL emission for retrospective dosimetryRadiation Measurements · 2025
- Calculation of the conversion coefficient from kerma to eye lens dose for computerized tomography head phantom.Radiation protection dosimetry · 2026
- From whole-body dosimetry to local dose assessment: a 3-y experience in interventional practice.Radiation protection dosimetry · 2026
- Dead-time Correction for High-count-rate WDS Analysis in Electron Probe Microanalysis.Microscopy and microanalysis : the official journal of Microscopy Society of America, Microbeam Analysis Society, Microscopical Society of Canada · 2026