OS13.7.A Confocal laser endomicroscopy outperforms white light and sodium fluorescein for tumour margin assessment in glioma surgery: the Florence experience
G Muscas, F Nozzoli, L Bini, A Della Puppa
Key points
From the paper's conclusion- In the recurrent HGG subgroup, a negative CLE result reduced the post-test probability of tumour infiltration from 65% to 31%, while negative NaF and WL findings provided no discriminative value above baseline prevalence.
Abstract
Background Identifying tumour infiltration at resection margins is the primary intraoperative challenge in glioma surgery. Confocal laser endomicroscopy (CLE) enables intraoperative real-time cellular imaging, but its diagnostic value relative to white-light microscopy (WL) and sodium fluorescein (NaF) has not been formally compared in a prospective setting across both primary and recurrent disease. Material and Methods In a prospective single-centre diagnostic accuracy study, 54 consecutive patients undergoing craniotomy for intracranial neoplasms — including primary tumours and reoperations for recurrent high-grade gliomas (HGG) — were enrolled. Resection margins were assessed sequentially with WL, NaF (YELLOW 560 filter), and CLE (ZEISS CONVIVO®) at the same anatomical site prior to biopsy for frozen section analysis. Results A total of 209 specimens were obtained; 133 (63.6%) were tumour-positive on frozen section, which served as the reference standard. The operating surgeon was blinded to pathological results; the neuropathologist was blinded to all imaging findings. CLE achieved the highest sensitivity (83.3%) and accuracy (73.9%), significantly outperforming NaF in accuracy and both WL and NaF in sensitivity and positive predictive value (all p ≤ 0.014). In multivariate mixed-effects analysis, CLE (OR 7.78, p = 0.002) and WL (OR 3.42, p = 0.035) were independent predictors of tumour infiltration; NaF was not significant after mutual adjustment. CLE identified histologically confirmed tumour in 65% of sites where both WL and NaF were simultaneously negative. The combined three-method model achieved AUC 0.933. CLE performance was robust across glioblastoma and lower-grade gliomas, but reduced in brain metastases (AUC 0.583). In the recurrent HGG subgroup, a negative CLE result reduced the post-test probability of tumour infiltration from 65% to 31%, while negative NaF and WL findings provided no discriminative value above baseline prevalence. Conclusion CLE proved superior to WL and NaF for the detection of tumour infiltration at the resection margin across primary neoplasms and recurrent gliomas. A negative CLE result provides clinically meaningful guidance in the recurrent setting, where fluorescence offers no incremental benefit. Sequential multimodal assessment combining all three methods maximises diagnostic accuracy and should be integrated into the neuro-oncological surgical workflow.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- OS13.6.A Delineation of the infiltration zone using intra-operative laserendomicroscopy in patients with glioblastomaNeuro-Oncology · 2026
- Real-time in vivo optical imaging of glioma with confocal laser endomicroscopy in a murine model: comparative analysis of performance of indocyanine green and fluorescein sodium.Frontiers in oncology · 2026
- OS13.5.A Prospective Multicenter Clinical Trial for Rapid Intraoperative Diagnosis and AI Augmented Evaluation of Brain Tumors Using ICG induced Confocal Laser EndomicroscopyNeuro-Oncology · 2026
- EP07.33 Integrating Confocal Laser Endomicroscopy and Label-Free Nanoplasmonic Biosensing for Intraoperative Glioblastoma Tissue DiscriminationNeuro-Oncology · 2026
- EP03.38 Temporal improvement in image acquisition quality and concordance using intraoperative confocal laser endomicroscopy in brain tumorsNeuro-Oncology · 2026
- Confocal Ex Vivo Margin Assessment in Transoral Laser Cordectomy: A Pilot Diagnostic Accuracy Study.The Laryngoscope · 2026