B-011 Evaluation of the Tosoh HLC-723GR01 HbA1c Analyzer Using Normal Hemoglobin Samples and in the Presence of Common Hemoglobin and Thalassemic Variants
Steven Hanson, Ian Gunsolus, Kuanysh Kabytaev
Key points
From the paper's conclusion- Conclusion The HLC-723GR01 demonstrated accurate HbA1c results for HbAA specimens and for samples containing common hemoglobin and thalassemic variants, with biases within predefined acceptance limits.
Abstract
Background HbA1c is used to aid in the diagnosis of diabetes, identify individuals at risk for diabetes, and monitor long-term glycemic control. Methods Venous whole blood collected in K2EDTA tubes from normal hemoglobin (HbAA) specimens (n=135) spanning 4.2%–16.6% HbA1c was tested as both whole blood and corresponding hemolysates (n=135). HbAA specimens were analyzed on the HLC-723G8 in triplicate and on the HLC-723GR01 as single measurements at three user-representative sites. Interference from common hemoglobin variants was assessed using HbAS (n=29), HbAC (n=25), HbAD (n=22), and HbAE (n=34), and from thalassemic variants using elevated HbA2 (n=33) and HbF (n=33), by comparing results to the HLC-723G8. Results For HbAA specimens, relative bias across HbA1c decision points ranged from 0.0% to 0.6% for both whole blood and hemolysates. Total error (bias plus imprecision) was =2.21% for whole blood and =2.49% for hemolysates across 5%–12% HbA1c. For variant specimens, relative bias ranged from -2.6% to 2.8%, meeting the acceptance criterion of =±7% at clinically relevant HbA1c levels (6.5% and 8.0%). Chromatographic separation supported consistent identification of the evaluated hemoglobin variants. Conclusion The HLC-723GR01 demonstrated accurate HbA1c results for HbAA specimens and for samples containing common hemoglobin and thalassemic variants, with biases within predefined acceptance limits.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-233 Haemoglobin Variants and Their Impact on HbA1c Measurement: Experience with Two Routine MethodsClinical Chemistry · 2026
- 70 Bio-Rad D-100 HbA1c Assay in Variant Populations: A Critical Analysis of Flagging and Reporting IssuesAmerican Journal of Clinical Pathology · 2026
- Evaluation of the Lifotronic H100 Plus analyzer for HbA1c determination.Scandinavian journal of clinical and laboratory investigation · 2026
- A-180 Measurement of hemoglobin A1c across three methodologies: A comparison in the presence and absence of hemoglobin variantsClinical Chemistry · 2026
- B-299 Analytical performance, variant detection and result reportability of the CAPILLARYS 3 TERA HbA1c assay relative to the Siemens Atellica CH930Clinical Chemistry · 2026
- Glycated Haemoglobin Versus Serum Fructosamine in Assessing Glycaemic Control Amongst Type 2 Diabetes Mellitus Patients With Haemoglobin Variants: A Systematic Review.BioMed research international · 2026