B-059 Precision and Bias Verification of HbA1c by HPLC Following CLSI EP15-A3 in a Hospital Laboratory in Manabí, Ecuador
Jorge Zambrano, Ana Alcívar, Ivon Howland
Key points
From the paper's conclusion- The aim of this study was to assess the precision and bias of the HbA1c method by HPLC following the CLSI EP15-A3 protocol in a hospital laboratory.
Abstract
Background Glycated Hemoglobin (HbA1c) is a key biomarker for the diagnosis and monitoring of diabetes mellitus, reflecting the average blood glucose over the past 2–3 months. The clinical reliability of the results depends on the proper analytical performance of the method, particularly in terms of precision and bias, which must be verified according to international standards and regulatory criteria. The aim of this study was to assess the precision and bias of the HbA1c method by HPLC following the CLSI EP15-A3 protocol in a hospital laboratory. Methods A Mindray H50P analyzer was used for HPLC to evaluate two levels of external control expressed as % NGSP, following the CLSI EP15-A3 protocol with a 5×5 design (n=25 per level). Results The results showed low intra-laboratory imprecision (=1.08%) and temporal stability of the method. The observed bias was minimal (0.18–0.53%) and remained well below the maximum allowable bias according to CLIA 2024 (8% of the assigned value: 0.45–0.91 units depending on the level), confirming high accuracy. Conclusion The HbA1c test by HPLC demonstrated high precision and minimal bias. These findings support the clinical implementation of the method, its integration into internal quality control, and ensure reproducible, reliable results that comply with clinical metrology standards
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of the Lifotronic H100 Plus analyzer for HbA1c determination.Scandinavian journal of clinical and laboratory investigation · 2026
- Comparison of Two HbA1c Measurement Methods and Their Analytical Performance: TINIA vs. Boronate Affinity HPLC.Clinical laboratory · 2026
- A-136 Improving HbA1c Result Reportability Using Dual-Method Verification: Bio-Rad D-100 HPLC and Siemens Healthineers DCA ImmunoassayClinical Chemistry · 2026
- B-005 Analytical Performance Evaluation of Two HbA1c Testing Methods on Roche cobas c513 AnalyzersClinical Chemistry · 2026
- Comparative analysis of enzymatic assay for HbA1c measurement with high performance liquid chromatography.Scientific reports · 2026
- Method comparison and agreement of a fluorescence immunoassay-based HbA1c test with a routine HPLC method.Diabetology international · 2026